Page values for "P6032"
Jump to navigation
Jump to search
"Products" values
1 row is stored for this pageManufacturer | Tektronix |
Model | P6032 |
Class | Probe |
Description | 850 MHz cathode-follower probe |
Series | 1S1 |
Introduced | 1962 |
Designers | Murlan Kaufman |
Discontinued | (?) |
"Specifications" values
9 rows are stored for this pageParameter | Bandwidth |
Value | ≥850 MHz |
Parameter | Rise time |
Value | 0.4 ns max. |
Parameter | Attenuation |
Value | 10× to 1000× in 1-2-5 steps with supplied attenuator heads, 3× without attenuator |
Parameter | Dynamic range |
Value | ±0.15 V × attenuation |
Parameter | Maximum input voltage |
Value | ±0.225 V × attenuation, de-rated for CW operation for 500× and 1000× attenuator heads |
Parameter | Input resistance |
Value | 10 MΩ with any attenuator head, 3 MΩ without attenuator head |
Parameter | Input capacitance |
Value | 3.6 pF (10× head) to 1.3 pF (1000× head), 7 pF without attenuator head |
Parameter | Output impedance |
Value | not specified, calibrated for 50 Ω load |
Parameter | DC offset |
Value | ≈0.5 V (not adjustable) |