[[File:p6042-complete.jpg|thumb|350px|right|P6042]] | [[File:p6042-complete.jpg|thumb|350px|right|P6042]] | ||
The '''Tektronix P6042''' is a DC to 50 MHz current probe. | The '''Tektronix P6042''' is a DC to 50 MHz current probe [[introduced in 1967]]. | ||
The sensor contains a current transformer for HF and a hall effect sensor for LF/DC components. | The sensor contains a current transformer for HF and a hall effect sensor for LF/DC components. | ||
Line 22: | Line 22: | ||
* [http://www.radiomuseum.org/r/tektronix_current_probe_dc_to_50_m.html Tek P6042 page @ radiomuseum.org] | * [http://www.radiomuseum.org/r/tektronix_current_probe_dc_to_50_m.html Tek P6042 page @ radiomuseum.org] | ||
* [http://www.pa4tim.nl/?p=3135 P6042 repair by Fred, PA4TIM] | * [http://www.pa4tim.nl/?p=3135 P6042 repair by Fred, PA4TIM] | ||
* [[Media:Service Scope 46 Oct 1968.pdf | Service Scope 46, October 1967]] | |||
==Pictures== | ==Pictures== | ||
Line 34: | Line 35: | ||
[[Category:Oscilloscope current probes]] | [[Category:Oscilloscope current probes]] | ||
[[Category:Introduced in 1967]] |
Revision as of 00:48, 4 October 2015
The Tektronix P6042 is a DC to 50 MHz current probe introduced in 1967.
The sensor contains a current transformer for HF and a hall effect sensor for LF/DC components. The amplifier combines DC/LF and HF signals.
Key Specifications
Bandwidth | DC to 50 MHz (-3 dB) |
---|---|
Rise time | 7 ns |
Ranges | 1 mA/Div to 1 A/Div (1-2-5) |
Output | to 50 Ω scope input at 50 mV/Div |
Maximum input | 10 A DC+peak AC in 5 mA to 1 A ranges, 0.5 A in 1 mA and 2 mA ranges |
Power | 90-136 or 180-272 V, 20 W |
Manuals
Links
- Tek P6042 page @ amplifier.cd
- Tek P6042 page @ barrytech.com
- Tek P6042 page @ radiomuseum.org
- P6042 repair by Fred, PA4TIM
- Service Scope 46, October 1967
Pictures
-
amplifier front
-
amplifier and sensor
-
sensor detail
-
rear view
-
catalog image