P6042

From TekWiki
Revision as of 14:32, 4 October 2015 by Soren (talk | contribs)
Jump to navigation Jump to search
P6042

The Tektronix P6042 is a DC to 50 MHz current probe introduced in 1967.

The sensor contains a current transformer for HF and a hall effect sensor for LF/DC components. The amplifier combines DC/LF and HF signals.

Key Specifications

Bandwidth DC to 50 MHz (-3 dB)
Rise time 7 ns
Ranges 1 mA/Div to 1 A/Div (1-2-5)
Output to 50 Ω scope input at 50 mV/Div
Maximum input 10 A DC+peak AC in 5 mA to 1 A ranges, 0.5 A in 1 mA and 2 mA ranges
Power 90-136 or 180-272 V, 20 W

Manuals

Links

Pictures