P6051

The Tektronix P6051 is a 1 GHz, ×5 active FET probe intended primarily for use with 50 Ω sampling instruments. It comes with additional ×4, ×20 and ×40 attenuator heads.

It has a GR-874 connector for the signal and a LEMO S-series connector to be powered by the scope or an 1101 PSU.

Key Specifications

Bandwidth 1000 MHz
Rise time 0.35 ns max.
Attenuation ×5, optional ×20, ×100 and ×200 with plug-on tips (010-0370-00, 010-0371-00, 010-0372-00)
Dynamic range ±2.5 V with ±5 V DC offset (× 1)
Maximum Input Voltage 200 V
Input impedance 1 MΩ // 2.8 pF (×1), 1 MΩ // 3.25 pF (×20), 1 MΩ // 2.0 pF (×100), 1 MΩ // 1.8 pF (×200)
Output impedance 50 Ω