P6057: Difference between revisions
Jump to navigation
Jump to search
No edit summary |
No edit summary |
||
(6 intermediate revisions by 2 users not shown) | |||
Line 1: | Line 1: | ||
The Tektronix P6057 | [[File:P6056.jpg|thumb|350px|right|P6056 probe (otherwise identical ×10 sister model)]] | ||
The '''Tektronix P6057''' is a 1.4 GHz ×100 probe for 50 Ω-input scopes, released along with the [[7A19]] amplifier in 1972. | |||
It has an Identify switch and readout contact. | |||
[[Category:50 | Its sister model [[P6056]] has ×10 attenuation. | ||
{{BeginSpecs}} | |||
{{Spec | Bandwidth | DC to 1.4 GHz }} | |||
{{Spec | Rise time | 250 ps }} | |||
{{Spec | Attenuation | 100× }} | |||
{{Spec | Input impedance | 5 kΩ // approx. 1 pF }} | |||
{{Spec | Scope input impedance| 50 Ω }} | |||
{{Spec | Maximum input Voltage| 50 V dc or RMS }} | |||
{{Spec | Features | Readout Pin, ID Button }} | |||
{{EndSpecs}} | |||
==Manual== | |||
* [http://w140.com/tek_P6056_P6057.pdf Tektronix P6056 and P6057 Manual (PDF)] | |||
<blockquote> | |||
<u>WARNING</u>: | |||
This high performance device has a fragile resistor assembly in the<br /> | |||
probe head, subject to failure in rough usage. Use with slip-on tip<br /> | |||
whenever possible. One replacement part is provided with the package.<br /> | |||
Contact nearest Field Office for additional parts (P/N 307-0330-00), if<br /> | |||
required. For replacement and Calibration Procedure, refer to Pg. 3 & 4<br /> | |||
of manual. | |||
</blockquote> | |||
[[Category:50 Ω Oscilloscope probes]] | |||
[[Category:Introduced in 1972]] |
Revision as of 21:29, 2 March 2021
The Tektronix P6057 is a 1.4 GHz ×100 probe for 50 Ω-input scopes, released along with the 7A19 amplifier in 1972.
It has an Identify switch and readout contact.
Its sister model P6056 has ×10 attenuation.
Key Specifications
Bandwidth | DC to 1.4 GHz |
---|---|
Rise time | 250 ps |
Attenuation | 100× |
Input impedance | 5 kΩ // approx. 1 pF |
Scope input impedance | 50 Ω |
Maximum input Voltage | 50 V dc or RMS |
Features | Readout Pin, ID Button |
Manual
WARNING:
This high performance device has a fragile resistor assembly in the
probe head, subject to failure in rough usage. Use with slip-on tip
whenever possible. One replacement part is provided with the package.
Contact nearest Field Office for additional parts (P/N 307-0330-00), if
required. For replacement and Calibration Procedure, refer to Pg. 3 & 4
of manual.