PG506/Repairs: Difference between revisions

m
FAILED PG506 and Repairs
m (Diagnosis of Failed Display section of PG506.)
m (FAILED PG506 and Repairs)
Line 18: Line 18:


WAVE FORMS Photographed on TDS460A.  See photos for scope settings as these can vary from photo to photo.
WAVE FORMS Photographed on TDS460A.  See photos for scope settings as these can vary from photo to photo.
During this repair, I found these parts defective.  The repairs were much more than a non-functional display.  The display was the last thing to get working.   
In the course of the repairs, I found these components bad in the unit.
Q435  151-0223-00  NPN
Q535  151-0223-00  NPN
U460  156-0067-00  OP amp  (741CN)
U200  156-0067-00  OP amp  (741CN)
U240  156-0067-00  OP amp  (741CN)
U610  156-0039-00  Dual J-K Flip-Flop
U480    156-0399-00 Opto Coupler
U255    156-0399-00 Opto Coupler
CR290  152-0141-02  (1N4152R)
CR291  152-0141-02  (1N4152R)
77