S-4: Difference between revisions

From TekWiki
Jump to navigation Jump to search
(Added link to local copy of instruction manual)
No edit summary
(One intermediate revision by the same user not shown)
Line 1: Line 1:
{{Plugin Sidebar 2 |
{{Plugin Sidebar
title=Tektronix S-4 |
|manufacturer=Tektronix
summary=Sampling Head |
|series=7000 and 3S series sampling heads
image=Tek-s-4.jpg |
|type=S-4
caption=S-4 head |
|summary=Sampling Head
series={{3S+7S}} |
|image=Tek-s-4.jpg
introduced=1968 |
|caption=S-4 head
discontinued=1990 |
|introduced=1968
manuals=
|discontinued=1990
|designers=George Frye
|manuals=
* [[Media:070-0896-01.pdf|Tektronix S-4 Instruction Manual, Revised OCT 1986 (OCR, PDF)]]
* [[Media:070-0896-01.pdf|Tektronix S-4 Instruction Manual, Revised OCT 1986 (OCR, PDF)]]
* [http://w140.com/tek_s4.pdf Tektronix S-4 Manual, copyright 1969 (PDF)]
* [http://w140.com/tek_s4.pdf Tektronix S-4 Manual, copyright 1969 (PDF)]
Line 44: Line 46:
It appears that a standard beam-lead diode may fit across the gap
It appears that a standard beam-lead diode may fit across the gap
but cleanly removing a failed diode without damaging the substrate would be quite difficult.
but cleanly removing a failed diode without damaging the substrate would be quite difficult.
==Prices==
1979: $1,435 (~$5,600 in 2022 Dollars)
According to an [[Media:Tek Schottky Diodes Memo rot.pdf|internal memo]], in 1979 annual sales were estimated at 150 units.


==Links==
==Links==
* [http://www.amplifier.cd/Test_Equipment/Tektronix/Tektronix_7000_series_special/S4.html S-4 page @ amplifier.cd]
* [http://www.amplifier.cd/Test_Equipment/Tektronix/Tektronix_7000_series_special/S4.html S-4 page @ amplifier.cd]
* [https://kh6htv.files.wordpress.com/2015/11/an-02a-oscopes.pdf James R. Andrews, ''Comparison of Ultra-Fast Rise Sampling Oscilloscopes''. Picosecond Pulse Labs App Note AN-2a, 1989]
* [https://kh6htv.files.wordpress.com/2015/11/an-02a-oscopes.pdf James R. Andrews, ''Comparison of Ultra-Fast Rise Sampling Oscilloscopes''. Picosecond Pulse Labs App Note AN-2a, 1989]


==Pictures==
==Pictures==

Revision as of 10:01, 13 March 2022

Manuals – Specifications – Links – Pictures

The Tektronix S-4 is a sampling head for 7000- and 3S-series samplers. It was designed by George Frye and introduced in 1968. It is the fastest of the S-series plug-in samplers.

Key Specifications

Rise time 25 ps (observed with S-50 or S-52, 35 ps)
Bandwidth 14.5 GHz
Input impedance 50 Ω (terminated SMA connector)
Input range 1 Vp-p (operating)
Maximum input ±5 V max. non-destructive
Noise < 5 mV
Features
  • trigger signal pick-off for internal triggering

Internals

The S-4 sampling gate is based upon a traveling wave trapped-charge transmission line in which the sampling window is set by the propagation time of a pulse edge through a thick-film transmission line. This technique requires only a sharp pulse edge rather than a precise pulse width, which is harder to generate. The sampling diodes are housed in a special coaxial connector that provides a high bandwidth signal path.

To disassemble the sampler hybrid, first remove it from the sampler board as per the manual. Remove the input connector using a 7/32" wrench and remove the 20 dB attenuator with small pliers. The ceramic board is held to the housing using roll pins that can be pressed out with a 0.030" pin punch. The hybrid has six diodes, each about 0.75mm square. The cathodes are glued to the gold substrate with conductive epoxy and the anodes are wire-bonded (twice) over a gap to the next step in the strobe line. It appears that a standard beam-lead diode may fit across the gap but cleanly removing a failed diode without damaging the substrate would be quite difficult.

Prices

1979: $1,435 (~$5,600 in 2022 Dollars)

According to an internal memo, in 1979 annual sales were estimated at 150 units.

Links

Pictures