S-52: Difference between revisions

From TekWiki
Jump to navigation Jump to search
No edit summary
No edit summary
Line 37: Line 37:
==Links==
==Links==
* [http://www.amplifier.cd/Test_Equipment/Tektronix/Tektronix_7000_series_special/S-52.html Tek S-52 page @ amplifier.cd]
* [http://www.amplifier.cd/Test_Equipment/Tektronix/Tektronix_7000_series_special/S-52.html Tek S-52 page @ amplifier.cd]
* http://www.barrytech.com/tektronix/tek7000/teks52.html
* [http://www.barrytech.com/tektronix/tek7000/teks52.html Tektronix S-52 @ barrytech.com]


==Pictures==
==Pictures==
Line 58: Line 58:
===Measurements===
===Measurements===
<gallery>
<gallery>
File:7s12-s52-incident.jpg | [[7S12]] displaying the incident pulse from an S-52 (nom. < 25 ps) through an [[S-6]] head (nom. < 30 ps).
File:7s12-self-portrait.jpg | TDR "Self Portrait" — pulse reflected on S-6 "through" path (approx. 120 ps one way)
File:7s12-s52-incident.jpg | 7S12 displaying the incident pulse from an S-52 (nom. < 25 ps) through an S-6 head (nom. < 30 ps). Displayed rise time ~35 ps confirms spec.
File:7s12-s52-pulse-1.jpg | Full pulse pattern of a 7S12/S-52 in real time (top trace) and sampled (bottom trace) shown simultaneously on [[7844]].  7S12 trace at slowest possible sweep.
File:7s12-s52-pulse-2.jpg | S-52 pulse in real time (top trace) and sampled (bottom trace) shown simultaneously on [[7844]].  Approximately equal time scales.
</gallery>
</gallery>




[[Category:7000 and 3S series sampling heads]]
[[Category:7000 and 3S series sampling heads]]

Revision as of 12:07, 27 April 2016

Template:Plugin Sidebar 2

The Tektronix S-52 is a pulse generator head compatible with S-series sampling units like the 3S2, 7S11, and 7S12.

It puts out a 200 mV pulse into a load of 50 Ω. The pulse has a rise time of 25 ps or less. The output is an SMA connector.

In addition to the main pulse output, the S-52 also produces a pre-trigger pulse 85 ns before the main pulse. The delay is generated with a digital counter, and jitter between the pre-trigger pulse and the main pulse is specified as less than 10 ps. The pre-trigger pulse uses a BSM connector.

Key Specifications

Rise time < 25 ps
Output amplitude 200 mV into 50 Ω (SMA)
Output impedance 50 Ω «reverse terminated» (i.e. having controlled source impedance)
Pre-trigger output 85 ns before main pulse (jitter < 10 ps), +1 V into 50 Ω (BSM)

Internals

The S-52 generates the main output pulse using a 50 mA low-capacitance tunnel diode with part number 153-0400-00.

The S-52 is similar to the S-50, but the S-50 has twice the pulse amplitude. The reason for this is that the S-52 has a 48 Ω resistor between the output tunnel diode and the output connector providing back termination, while the S-50 has no such resistor. Although the S-52 produces a smaller pulse, its output impedance is much better controlled than that of the S-50, and the output resistor may also provide some limited protection for the output tunnel diode.