SG503: Difference between revisions

From TekWiki
Jump to navigation Jump to search
No edit summary
No edit summary
Line 6: Line 6:


Nine overlapping ranges cover the frequency band from 250 kHz to 250 MHz,with an additional range reserved for a 50 kHz reference frequency.
Nine overlapping ranges cover the frequency band from 250 kHz to 250 MHz,with an additional range reserved for a 50 kHz reference frequency.
The SG503 shows the selected frequency on a 3-digit LED counter.


{{BeginSpecs}}
{{BeginSpecs}}
{{Spec | Frequency |   }}
{{Spec | Frequency | 50 kHz, 250 kHz to 250  MHz in nine ranges  }}
{{Spec | Amplitude |   }}
{{Spec | Frequency accuracy |  Within ±0.7% of one count of the least significant displayed digit for the indicated frequency }}
{{Spec | Amplitude | 5 mV<sub>p-p</sub> to 5.5 V<sub>p-p</sub> in three decade ranges (terminated into a 50 Ω load)  }}
{{Spec | Harmonics | 2<sup>nd</sup> harmonic ≥35 dB down from fundamental, 3<sup>rd</sup> and higher harmonics ≥40 dB down }}
{{EndSpecs}}
{{EndSpecs}}



Revision as of 02:35, 5 March 2017

Tektronix SG503
250 MHz leveled sinewave generator
File:Sg503.jpg
Tektronix SG503

Produced from 1974 to 1993

Manuals
(All manuals in PDF format unless noted otherwise)
Manuals – Specifications – Links – Pictures

The Tektronix SG503 is a 250 MHz leveled sinewave generator plug-in for the TM500 system.

It was primarily intended to be used as an oscilloscope calibration device for measuring bandwidths up to 250 MHz. Unlike later instruments, it does not use a leveling head but was designed and calibrated with high quality coaxial cable (P/N 012-0482-00) to operate as a closely matched system when terminated into a 50 Ω load.

Nine overlapping ranges cover the frequency band from 250 kHz to 250 MHz,with an additional range reserved for a 50 kHz reference frequency.

The SG503 shows the selected frequency on a 3-digit LED counter.

Key Specifications

Frequency 50 kHz, 250 kHz to 250  MHz in nine ranges
Frequency accuracy Within ±0.7% of one count of the least significant displayed digit for the indicated frequency
Amplitude 5 mVp-p to 5.5 Vp-p in three decade ranges (terminated into a 50 Ω load)
Harmonics 2nd harmonic ≥35 dB down from fundamental, 3rd and higher harmonics ≥40 dB down

Pictures