[[Image:P6046.jpg|350px|thumb|right|Tektronix P6046 Differential Probe]]
{{Probe Sidebar
|manufacturer=Tektronix
|model=P6046
|summary=100 MHz differential probe
|image=Tek P6046 catalog 2.jpg
|caption=Tektronix P6046 Differential Probe and 015-0106-00 Amplifier
|for=50 Ω scope inputs
|introduced=1968
|discontinued=1992
|designers=Glenn Bateman
|manuals=
There are three versions of the manual:
* [[Media:070-0756-00.pdf|070-0756-00 - 1968]] − assumes [[1A5]] for absolute gain calibration
* [[Media:070-7129-00.pdf|070-7129-00 - 1988]]
* [[Media:070-7129-01.pdf|070-7129-01 - 1992]]


The ''Tektronix P6046'' is a differential probe and amplifier.
* [[Media:P6046 Voltage and Waveform Information.pdf|P6046 Voltage and Waveform Information]]
It is mainly intended for use with a [[1A5]] and attaches through a special multi-pin connector.
* [[Media:P6046 Schematics Rev. C June 1977 11x17.pdf|P6046 Schematics Rev. C June 1977]]


With a 1A5, it has a 45 MHz bandwidth.  The raw bandwidth of the probe and amplifier is 100 MHz.
* [[Media:Tek p6046 eis signature page.pdf|P6046 Engineering Instrument Specification Signature Page]]
* [[Media:Tek p6046 fcp march 1968.pdf|P6046 Factory Calibration Procedure, March 1968]]
}}
The '''Tektronix P6046''' is a 100 MHz differential probe oscilloscope probe [[introduced in 1968]] and discontinued sometime after 1992.
It was designed by [[Glenn Bateman]].


==Manuals==
It was initially intended for use with a [[1A5|1A5 plug-in]], to which it attaches through a 9-pin [[Amphenol 165 series connector]].


* [http://w140.com/Tektronix_P6046_Instruction_Manual.pdf P6046 Manual (no schematics, PDF)]
A separate amplifier (015-0106-00) was available as an option to use the probe on a 50 Ω input scope set at 10 mV/Div.
* [http://w140.com/Tek_P6046_Differential_Probe-Operation_Manual.pdf P6046 Detailed Manual (no schematics, PDF)]
 
* [http://w140.com/P6046_8.5x11.pdf P6046 Voltage and Waveform Information (PDF)]
An additional ×10 attenuator (010-0361-00) can be plugged onto the P6046's inputs.
* [http://w140.com/P6046_17x11.pdf P6046 Schematics (PDF)]
 
Note that the probe itself acts as either a 1× or 10× attenuation probe dependent on a signal voltage on pin D of the connector. 
When used with the amplifier, 10× mode is used in the 200, 100, and 50 mV/div settings and 1× mode for 20, 10, 5, 2, and 1 mV/div.
 
{{BeginSpecs}}
{{Spec | Bandwidth | 45 MHz with 1A5; 100 MHz with amplifier; AC low frequency cut-off 20 Hz (2 Hz with ×10 head) }}
{{Spec | Input impedance | 1 MΩ // 10 pF (with ×10 head, 10 MΩ // 3 pF) }}
{{Spec | Common-mode range | ±5 V DC+peak AC up to 10 MHz, decreasing to ±2 V at 50 MHz }}
{{Spec | Maximum Allowable Input | 25 V total DC+peak AC, 25 V total difference between + and − input tips }}
 
{{EndSpecs}}
 
==Operation==
 
The two probe tips are (in DC mode) directly connected to the gates of a pair of JFETs. 
 
Use the shorting plugs on the tips when not in use:
<blockquote>Ground the probe!  Seriously.  Sticker the probe "Ground Me!"  I also recommend stickering the probe amplifier with the ATTEN BAL setup procedure from section 2-6 of the manual.</blockquote>
 
==Links==
* [http://www.i9t.net/p6046/p6046.html P6046 @ iceNINE Tech]
* [https://www.amplifier.cd/Test_Equipment/Tektronix/Tektronix_other/P6046.htm P6046 @ amplifier.cd]
{{Documents|Link=P6046}}


==Pictures==
==Pictures==
<gallery>
<gallery>
File:Tek p6046 internal1.jpg
P6046_1.jpg | P6046
File:Tek p6046 internal2.jpg
P6046_2.jpg | P6046
File:Tek p6046 coupling.jpg
Tek p6046 internal1.jpg
Tek p6046 internal2.jpg
Tek p6046 coupling.jpg
015-0106-00.jpg | 015-0106-00 amplifier and power supply
</gallery>
</gallery>


[[Category:Active Oscilloscope Probes]]
[[Category:Differential oscilloscope probes]]

Latest revision as of 06:14, 28 January 2024

The Tektronix P6046 is a 100 MHz differential probe oscilloscope probe introduced in 1968 and discontinued sometime after 1992. It was designed by Glenn Bateman.

Tektronix P6046
100 MHz differential probe
Tektronix P6046 Differential Probe and 015-0106-00 Amplifier

Compatible with 50 Ω scope inputs

Produced from 1968 to 1992

Manuals


It was initially intended for use with a 1A5 plug-in, to which it attaches through a 9-pin Amphenol 165 series connector.

A separate amplifier (015-0106-00) was available as an option to use the probe on a 50 Ω input scope set at 10 mV/Div.

An additional ×10 attenuator (010-0361-00) can be plugged onto the P6046's inputs.

Note that the probe itself acts as either a 1× or 10× attenuation probe dependent on a signal voltage on pin D of the connector. When used with the amplifier, 10× mode is used in the 200, 100, and 50 mV/div settings and 1× mode for 20, 10, 5, 2, and 1 mV/div.

Key Specifications

Bandwidth 45 MHz with 1A5; 100 MHz with amplifier; AC low frequency cut-off 20 Hz (2 Hz with ×10 head)
Input impedance 1 MΩ // 10 pF (with ×10 head, 10 MΩ // 3 pF)
Common-mode range ±5 V DC+peak AC up to 10 MHz, decreasing to ±2 V at 50 MHz
Maximum Allowable Input 25 V total DC+peak AC, 25 V total difference between + and − input tips

Operation

The two probe tips are (in DC mode) directly connected to the gates of a pair of JFETs.

Use the shorting plugs on the tips when not in use:

Ground the probe! Seriously. Sticker the probe "Ground Me!" I also recommend stickering the probe amplifier with the ATTEN BAL setup procedure from section 2-6 of the manual.

Links

Documents Referencing P6046

Document Class Title Authors Year Links
Service Scope 46 Oct 1967.pdf Article P6046 DC-to-100 MHz Differential Probe and Amplifier Glenn Bateman 1967
Tekscope 1972 V4 N5 Sep 1972.pdf Article A Practical Approach to Differential Amplifiers and Measurements Fred Beckett 1972
Tekscope 1972 V4 N6 Nov 1972.pdf Article Differential Amplifiers and Measurements, Part 2 Fred Beckett 1972

Pictures