Page values for "Telequipment Oscilloscope Calibrator C3"

Jump to navigation Jump to search

"_pageData" values

1 row is stored for this page
FieldField typeValue
_creationDateDatetime2020-01-17 6:53:20 PM
_modificationDateDatetime2022-02-20 9:47:55 PM
_categoriesList of String, delimiter: |CRT based solid-state scopes Introduced in (?) Plugin scopes Telequipment
_isRedirectBooleanNo
_pageIDInteger23,035
_pageNamePageTelequipment Oscilloscope Calibrator C3
_pageTitleString

Telequipment Oscilloscope Calibrator C3

_pageNamespaceInteger0

"Products" values

1 row is stored for this page
FieldField typeValue
ManufacturerPageTelequipment
ModelStringOscilloscope Calibrator C3
ClassStringCalibrator
DescriptionWikitext string

Oscilloscope Calibrator

SeriesPage
IntroducedText(?)
DesignersList of Page, delimiter: ;
DiscontinuedText(?)

"Specifications" values

9 rows are stored for this page
FieldField typeValue
ParameterStringMarkers
ValueWikitext

100 ns to 5 s, 1–2–5; pulse amplitude ~1 V; pulse width ~2% of period; accuracy ±0.01%

FieldField typeValue
ParameterStringAmplitude Calibrator
ValueWikitext

10 kHz / 1 kHz / 50 Hz / 10 Hz / 1 Hz square wave, or DC; amplitude 100 V to 0.025 V and 100 mV to 0.025 mV, 1–2–5; rise time ≤ 100 ns; 50 Hz output has rise time of 400 μs. Accuracy ±0.2% when loaded with 1 MΩ except 0.05 V/mV ±0.3% and 0.025 V/mV ±0.35%; output impedance varies from 5 kΩ at 50 V and 50 mV down to 75 Ω at ≤0.5 V and 0.5 mV.

FieldField typeValue
ParameterStringFast Rise Signal
ValueWikitext

Square waveform switched at 100 kHz, 1 MHz, 10 MHz; rise time ≤ 1 ns; output 0.2 V max. into 50 Ω

FieldField typeValue
ParameterStringTelevision Waveform
ValueWikitext

625-line signal, +ve and —ve modulation; amplitude 1.6 Vp-p; output impedance 75 Ω; -ve going output is DC coupled from ground, +ve going output is AC coupled

FieldField typeValue
ParameterStringTV signal modulation
ValueWikitext

Black, White, or square wave at 62.5 kHz / 250 kHz / 1 MHz / 5 MHz; Modulation can be switched on and off at one-second rate.

FieldField typeValue
ParameterStringAuxiliary Outputs
ValueWikitext

Available on 4 mm sockets at the side; square waves at 10 MHz, 1 MHz, 100 kHz, 10 kHz, 1 kHz, 100 Hz, 10 Hz, 1 Hz; amplitude ~2.5 V, output impedance ~1 kΩ

FieldField typeValue
ParameterStringExt Clock
ValueWikitext

External clock frequencies of 2 V peak to peak amplitude may be fed to the 10 MHz Auxiliary Output socket

FieldField typeValue
ParameterStringPower
ValueWikitext

100, 110, 120, 200, 220, 240 V, 48 to 440 Hz, 13 VA.

FieldField typeValue
ParameterStringSize
ValueWikitext

107 mm × 228 mm × 285 mm (H×W×D)