576/Repairs: Difference between revisions

no edit summary
No edit summary
 
(2 intermediate revisions by 2 users not shown)
Line 82: Line 82:
RESULT: -75V drifts badly as the machine is turned on and heats up (High @ -73.5V when turned on and drifting slowly to -75.05V when warmed up.).  
RESULT: -75V drifts badly as the machine is turned on and heats up (High @ -73.5V when turned on and drifting slowly to -75.05V when warmed up.).  
   
   
Found shorted D83 -  1N4152 not aware of symptoms related to this defect.
Found shorted D83 -  1N4152 not aware of symptoms related to this defect.  This would have affected the DAC Function.  See Page 3-9 for Circuit Description.
      
      
Replaced D708 with proper 9V TC ZENER, -75V Supply is now rock solid @-75.015V, no "drift" as the machine is turned on and as it heats up.  
Replaced D708 with proper 9V TC ZENER, -75V Supply is now rock solid @-75.015V, no "drift" as the machine is turned on and as it heats up.  
Line 98: Line 98:
Replaced K101 K102 and K103 relays.  
Replaced K101 K102 and K103 relays.  
No symptoms or malfunctions noted.
No symptoms or malfunctions noted.
==Symptom: Complete Unit shut down #2==
Bad HV transformer.  Unit #2 quit working after a short time of operation.  Very low S/N B010169 unit is very old and very early production model.  Had brown epoxy potted HV XFMR.  No luck with finding a  functional new or used part.  Found a manufacturer who was willing to make a small batch of these transformers, contracted for a batch to be made. Replaced old part with new one, unit works again.  There is copious discussion on the TEKScopes Group regarding this repair.  250+ hours of operation proves this part is durable.  I have very few of these transformers remaining available for sale. (Contact Me OFF-LIST for details.) 
[[Category:Instrument repair reports]]
77

edits