[[File:Tek p6000.jpg|thumb|300px|right]]
{{Probe Sidebar
The '''Tektronix P6000''' is a 30 MHz,  ×10 passive voltage probe designed for use with  
|manufacturer=Tektronix
oscilloscopes that have a 1 MΩ // 20 to 50 pF input impedance and a [[UHF connector]].
|model=P6000
|summary=30 MHz ×10 Passive Probe
|image=Tektronix_P6000_Probe_2.jpg
|caption=Tektronix P6000
|for=1 MΩ scope inputs
|series=
|introduced=(?)
|discontinued=(?)
|manuals=
* [[Media:Tek_p6000_old.pdf|Tektronix P6000 Manual]] (older)
* [[Media:Tek_p6000_new.pdf|Tektronix P6000 Manual]] (newer)
* [[Media:Tek_p6000-to-p6005_parts.pdf|Tek P6000 to P6005 Parts]]
}}
The '''Tektronix P6000''' is a 30 MHz, ×10 passive voltage probe designed for use with oscilloscopes that have a 1 MΩ // 20 to 50 pF input impedance and a [[UHF connector]].
The [[P6003]] is a variant of the [[P6000]] that has a [[BNC connector]] instead of a UHF connector.
The [[P6003]] is a variant of the [[P6000]] that has a [[BNC connector]] instead of a UHF connector.
The compensation is done by changing the length of the probe body itself, which is achieved by turning the front part after loosening the finger stop which also serves as a twist-lock.


Probe input impedance is 10 MΩ // 11.5 to 14.5 pF
{{BeginSpecs}}
Maximum voltage is 1200 V.
{{Spec | Input impedance   | 10 MΩ // 11.5–14.5 pF }}
The connection with the scope is with
{{Spec | Compensation range | 1 MΩ // 20–50 pF }}
 
{{Spec | Maximum input      | 1200 V DC or AC peak}}
The P6000 was made with four different cable lengths:
{{Spec | Cable |
* 3.5 feet, part number 010-0020-00
* 3.5 feet, part number 010-0020-00
* 6 feet, part number 010-0030-00
* 6 feet, part number 010-0030-00
* 9 feet, part number 010-0035-00
* 9 feet, part number 010-0035-00
* 12 feet, part number 010-0041-00
* 12 feet, part number 010-0041-00
}}
{{EndSpecs}}


==Manuals==
==Pictures==
 
<gallery>
* [http://w140.com/tek_p6000_old.pdf Tektronix P6000 Manual (older, PDF)]
Tektronix_P6000_Probe_2.jpg
* [http://w140.com/tek_p6000_new.pdf Tektronix P6000 Manual (newer, PDF)]
Tektronix_P6000_probe.jpg
* [http://w140.com/tek_p6000-to-p6005_parts.pdf Tek P6000 to P6005 Parts (PDF)]
Tektronix_P6000_head.jpg
 
Tektronix_P6000_head_2.jpg
Tektronix P6000 compensation.JPG
Tektronix_P6000_UHF_connector.jpg
</gallery>


[[Category:1 MOhm Oscilloscope Probes]]
[[Category:1 Oscilloscope probes]]
[[Category:6-32 probes]]

Latest revision as of 00:13, 4 July 2023

Tektronix P6000
30 MHz ×10 Passive Probe
Tektronix P6000

Compatible with 1 MΩ scope inputs

Produced from (?) to (?)

Manuals


Manuals – Specifications – Links – Pictures

The Tektronix P6000 is a 30 MHz, ×10 passive voltage probe designed for use with oscilloscopes that have a 1 MΩ // 20 to 50 pF input impedance and a UHF connector. The P6003 is a variant of the P6000 that has a BNC connector instead of a UHF connector. The compensation is done by changing the length of the probe body itself, which is achieved by turning the front part after loosening the finger stop which also serves as a twist-lock.

Key Specifications

Input impedance 10 MΩ // 11.5–14.5 pF
Compensation range 1 MΩ // 20–50 pF
Maximum input 1200 V DC or AC peak
Cable
  • 3.5 feet, part number 010-0020-00
  • 6 feet, part number 010-0030-00
  • 9 feet, part number 010-0035-00
  • 12 feet, part number 010-0041-00