067-0502-00: Difference between revisions
No edit summary |
m (Capitolize sidebar summary) |
||
(19 intermediate revisions by 4 users not shown) | |||
Line 1: | Line 1: | ||
The Tektronix 067-0502-00 is an instrument designed for calibrating oscilloscope calibrators. | {{Instrument Sidebar | ||
It is very similar to the [[067-0502-01]]. | |manufacturer=Tektronix | ||
|model=067-0502-00 | |||
|class=Signal generator | |||
|series= | |||
|summary=Amplitude Calibrator | |||
|image=067-0502-00 front.jpg | |||
|caption=Tektronix 067-0502-00 | |||
|introduced=(?) | |||
|discontinued=(?) | |||
|designers= | |||
|manuals= | |||
* [[Media:062-0808-00.pdf|Tektronix 067-0502-00 Manual]] (PDF) | |||
* [[Media:Tek 067-0502-00.pdf|Tektronix 067-0502-00 Manual]] (PDF, OCR) | |||
}} | |||
The '''Tektronix 067-0502-00''' is an instrument designed for calibrating oscilloscope calibrators. | |||
It is very similar to the [[067-0502-01]]. | |||
The 067-0502-00 replaced the [[067-0077-00]]. | |||
{{BeginSpecs}} | |||
{{Spec | Output | DC+, DC-, Squarewave}} | |||
{{Spec | Amplitude range | 0.2 mV to 100 V in 1−2−5 sequence, 5 mA in current loop}} | |||
{{Spec | Amplitude accuracy | Within 0.25% into 1 MΩ load}} | |||
{{Spec | Squarewave repetition rate | 1 kHz within 20% }} | |||
{{Spec | Maximum input voltage | 100 V }} | |||
{{Spec | 100x Amplifier gain accuracy | ±1% }} | |||
{{Spec | 100x Amplifier risetime | < 35 μs }} | |||
{{Spec | 100x Amplifier output range | ≥ ±20 V, ±5 mA }} | |||
{{Spec | 100x Amplifier Output resistance | approx. 250 Ω}} | |||
{{EndSpecs}} | |||
==Comparison with 067-0502-01== | |||
Sherman Foy contributed the following information: | Sherman Foy contributed the following information: | ||
<blockquote> | <blockquote> | ||
TekWiki asks the question: | TekWiki asks the question: What is the difference? I have both units and have done an extensive review. | ||
Looking at the front panels [TNX 4 pix] of the –00 & the –01, it seems that the –01 dropped the × 100 Amplifier section. | |||
The front panel is re-laid out, but what remains seems to be identical to the corresponding sections from the –00. | |||
In other words, the –00 provides a superset of the functionality of the -01. | |||
Note that the –00 transformer [T310] has 6 windings and the –01 transformer [also T310] has 5 windings. | |||
Some currents are different, as indicated. | |||
I have made my own notes for the project of working on the two. | |||
Using the schematics for organization, I arrived at the following table. | |||
I am using a re-organization of the sheets, the block diagram of the –00 and this table: | |||
{| class="wikitable" | {| class="wikitable" | ||
|- | |- | ||
| Section | ! scope="col"| Section | ||
| -00 | ! scope="col"| -00 | ||
| -01 | ! scope="col"| -01 | ||
| Sheet in -00 | ! scope="col"| Sheet in -00 | ||
| Sheet in -01 | ! scope="col"| Sheet in -01 | ||
| Sheet in proposed consolidated drawing set | ! scope="col"| Sheet in proposed consolidated drawing set | ||
|- | |- | ||
| Square Wave calibrator | | Square Wave calibrator | ||
Line 29: | Line 63: | ||
| [s/b 1] | | [s/b 1] | ||
|- | |- | ||
| | | × 100 Amplifier | ||
| Used in -00 | | Used in -00 | ||
| | | Deleted in -01 | ||
| 2 | | 2 | ||
| -- | | -- | ||
Line 51: | Line 85: | ||
|- | |- | ||
| High voltage, B+ | | High voltage, B+ | ||
| | | 285 V<sub>RMS</sub>, 35 mA 300 V<sub>DC</sub>, 225 V<sub>DC</sub> & 100 V<sub>DC</sub> | ||
| | | 125 V<sub>RMS</sub>, 2 mA 100 V<sub>DC</sub> only | ||
| 3 | | 3 | ||
| 2 | | 2 | ||
| 3A/3B | | 3A/3B | ||
|- | |- | ||
| 6. | | 6.3 V<sub>AC</sub> (Chopper) | ||
| Used in -00 | | Used in -00 | ||
| Used in -01, noted as 1. | | Used in -01, noted as 1.84 A | ||
| 3 | | 3 | ||
| 2 | | 2 | ||
| 3A/3B | | 3A/3B | ||
|- | |- | ||
| | | 100 V<sub>RMS</sub> (-100 V<sub>DC</sub>) | ||
| Used in -00, | | Used in -00, 40 mA | ||
| Used in -01, | | Used in -01, 31 mA | ||
| 3 | | 3 | ||
| 2 | | 2 | ||
| 3A/3B | | 3A/3B | ||
|- | |- | ||
| 6. | | 6.3 V<sub>AC</sub> (MV 6DJ8) | ||
| Used in -00 | | Used in -00 | ||
| Used in -01 | | Used in -01 | ||
Line 78: | Line 112: | ||
| 3A/3B | | 3A/3B | ||
|- | |- | ||
| | | 330 V<sub>RMS</sub> (Floating Supply) | ||
| Used in -00 | | Used in -00 | ||
| Used in -01 | | Used in -01 | ||
Line 85: | Line 119: | ||
| 3A/3B | | 3A/3B | ||
|- | |- | ||
| | | 20 V<sub>RMS</sub> (× 100 Filaments) | ||
| Used in -00 | | Used in -00 | ||
| Not used in –01 T310 | | Not used in –01 T310 | ||
Line 99: | Line 133: | ||
| [s/b 4] | | [s/b 4] | ||
|} | |} | ||
</blockquote> | |||
==Pictures== | |||
<gallery> | <gallery> | ||
067-0502-00 front.jpg | |||
067-0502-00.JPG | |||
</gallery> | </gallery> | ||
[[Category:Calibration fixtures]] | [[Category:Calibration fixtures]] | ||
[[Category:Amplitude Calibrators]] |
Latest revision as of 19:42, 15 November 2023
The Tektronix 067-0502-00 is an instrument designed for calibrating oscilloscope calibrators. It is very similar to the 067-0502-01. The 067-0502-00 replaced the 067-0077-00.
Key Specifications
Output | DC+, DC-, Squarewave |
---|---|
Amplitude range | 0.2 mV to 100 V in 1−2−5 sequence, 5 mA in current loop |
Amplitude accuracy | Within 0.25% into 1 MΩ load |
Squarewave repetition rate | 1 kHz within 20% |
Maximum input voltage | 100 V |
100x Amplifier gain accuracy | ±1% |
100x Amplifier risetime | < 35 μs |
100x Amplifier output range | ≥ ±20 V, ±5 mA |
100x Amplifier Output resistance | approx. 250 Ω |
Comparison with 067-0502-01
Sherman Foy contributed the following information:
TekWiki asks the question: What is the difference? I have both units and have done an extensive review.
Looking at the front panels [TNX 4 pix] of the –00 & the –01, it seems that the –01 dropped the × 100 Amplifier section. The front panel is re-laid out, but what remains seems to be identical to the corresponding sections from the –00. In other words, the –00 provides a superset of the functionality of the -01.
Note that the –00 transformer [T310] has 6 windings and the –01 transformer [also T310] has 5 windings. Some currents are different, as indicated.
I have made my own notes for the project of working on the two. Using the schematics for organization, I arrived at the following table. I am using a re-organization of the sheets, the block diagram of the –00 and this table:
Section -00 -01 Sheet in -00 Sheet in -01 Sheet in proposed consolidated drawing set Square Wave calibrator Used in both Used in both 1 1 [s/b 1] × 100 Amplifier Used in -00 Deleted in -01 2 -- [s/b 2] Power Supply -00 has more sections Than –01 3 2 [s/b 3A for -00 & 3B for -01] T310 P/N 120-0442-00 P/N 120-0615-00 3 2 3A/3B High voltage, B+ 285 VRMS, 35 mA 300 VDC, 225 VDC & 100 VDC 125 VRMS, 2 mA 100 VDC only 3 2 3A/3B 6.3 VAC (Chopper) Used in -00 Used in -01, noted as 1.84 A 3 2 3A/3B 100 VRMS (-100 VDC) Used in -00, 40 mA Used in -01, 31 mA 3 2 3A/3B 6.3 VAC (MV 6DJ8) Used in -00 Used in -01 3 2 3A/3B 330 VRMS (Floating Supply) Used in -00 Used in -01 3 2 3A/3B 20 VRMS (× 100 Filaments) Used in -00 Not used in –01 T310 3 2 3A/3B Floating power supply Used in both; drives Square Wave MV Used in both; drives Square Wave MV 4 3 [s/b 4]