2714: Difference between revisions

From TekWiki
Jump to navigation Jump to search
No edit summary
No edit summary
 
(14 intermediate revisions by 3 users not shown)
Line 1: Line 1:
{{Oscilloscope Sidebar |
{{Instrument Sidebar  
title=Tektronix 2714 |
|class=Spectrum Analyzer
summary=Spectrum Analyzer 9 kHz to 1.8 GHz |
|manufacturer=Tektronix  
image= Tek_2714_Waveform_image.jpg |
|model=2714  
caption=Tektronix 2714 |
|summary=1.8 GHz Spectrum Analyzer  
introduced=1993 |
|image= Tek_2714_Waveform_image.jpg  
discontinued=2000 |
|caption=Tektronix 2714  
manuals=
|introduced=1993  
Need 2714 Service Manual
|discontinued=2000  
* [[Media:070-8532-00.pdf | Tektronix 2714 User Manual (PDF)]]
|manuals=
* [[Media:070-8533-04.pdf | Tektronix 2714 & 2715 Programmer Manual (PDF)]]
* [[Media:070-8532-00.pdf|Tektronix 2714 User Manual]]
* [[Media:070-8535-06.pdf|Tektronix 2714/2715 Reference]]
* [[Media:070-8533-05.pdf|Tektronix 2714/2715 Programmer Manual -05]]
* [[Media:070-8533-04.pdf|Tektronix 2714/2715 Programmer Manual -04]]
* [[Media:2EW-10598-0.pdf| Cable TV Measurements using the 2714/1715 Spectrum Analyzers]]
* Service manual needed [[Category:Manual needed]]
}}
}}
The '''Tektronix 2714''' is a CATV spectrum analyzer with a frequency range of 9 kHz to 1.8 GHz.


The '''Tektronix 2714''' is a CATV spectrum analyzer with a frequency range of 9 kHz to 1.8 GHz in coax.  The minimum resolution bandwidth is 300 Hz, and provides measurement resolution proportional to the frequency accuracy or span.  With non-volatile memory (NVRAM) at 124 kB can save up to 108 displays of waveforms and markers, and up to 36 front panel setups, large user-defined key programs, and antenna factor tables.
The 2714 Cable TV Spectrum Analyzer supports CATV-specific functions like automatic positioning of visual aural carriers, visual and aural carrier levels and frequencies for selected channels, survey of system visual and aural carrier levels and frequencies, depth of modulation, aural carrier deviation measurement, visual carrier-to-noise measurement, and visual to carrier amplitude and frequency difference measurements.
 
The single and delta markers provide direct readout of frequency and amplitude information of any point along any displayed trace.  Or, get the relative (delta) frequency and amplitude information between any two points along any displayed trace or between traces.
 
The 2714 Cable TV Spectrum Analyzer has features to support the measurements necessary to maintain a functional CATV system like: Automatic positioning of visual aural carriers, Visual and aural carrier levels and frequencies for selected channels, Survey of system visual and aural carrier levels and frequencies, Depth of modulation, Aural carrier deviation measurement, Visual carrier-to-noise measurement, and Visual to carrier amplitude and frequency difference measurements.
 
Additional features include Digital Storage Display (selectable acquisition modes of positive peak only, positive/negative peak, Max hold, Waterfall, and an analog sweep when digital storage is off), Ensemble Averaging, Internal Preamp, Internal Calibrator (accessed through Input Menu), Alternate Reference Level Units (dBm, dBmV, dBV, dBµV, dBµW, and dBµV/m), User-Definable Power-on Status, Center Measure, Graticule Illumination (for CRT photography), and Direct Plot/Print (Printers via Opt. 03 GPIB, or RS-232-C interface via Opt. 08).
 
Front panel inputs/outputs are: RF INPUT [[BNC connector]] (75Ω, +20 dBm Max, 100 V DC blocking capacitor); and OPTION [[Unknown connector]] (Service manual only indicates that the front panel position is for a '''Future Option'''.  No internal tracking generators were available like with the other 2710, 2711, and 2712).
 
Rear panel inputs/outputs are: J101 [[SMA connector]] (Female SMA installed for Opt. 15), J102 EXT TRIG [[BNC connector]] (10 kΩ, 50V (DC + peak AC)), J103 [[DB9 connector]] Female (Not RS-232 compatible.  Accessory connector: Pin 1 Video Input, Pin 2 Ground, Pin 3 Video Output, Pin 6 Sweep Gate, Pin 7 Sweep Output, all others unused), J104 Opt. 03: [[GPIB connector]] (IEEE Std 488 connector) or Opt. 08: [[RS-232 interface]] (Digital Communication Port.).


{{BeginSpecs}}
{{BeginSpecs}}
{{Spec | Frequency Range | 9 kHz to 1.8 GHz; Preamp limited to 600 MHz }}
{{Spec | Frequency | 9 kHz to 1.8 GHz; Preamp limited to 600 MHz }}
{{Spec | Frequency Span Range (plus 0 Hz and MAX: 180 MHz/div) | 1 kHz/div to 100 MHz/div in 1, 2, 5 sequence }}
{{Spec | Frequency Span | 1 kHz/div to 100 MHz/div in 1, 2, 5 sequence (plus 0 Hz and MAX: 180 MHz/div) }}
{{Spec | Resolution Bandwidth Range | -6 dB bandwidth selections are 300 Hz, 1 kHz, 3 kHz, 10 kHz, 30 kHz, 100 kHz, 300 kHz, 1 MHz and 5 MHz }}
{{Spec | Resolution Bandwidth | –6 dB bandwidth selections are 300 Hz, 1 kHz, 3 kHz, 10 kHz, 30 kHz, 100 kHz, 300 kHz, 1 MHz and 5 MHz }}
{{Spec | RF Input impedance | 75Ω }}
{{Spec | RF Input impedance | 75 Ω }}
{{Spec | Maximum Safe Input Power | +70 dBmV (0.1 W or 2.2 V) continuous peak, ≤100 V DC blocking capacitor }}
{{Spec | Maximum Safe Input Power | +70 dBmV (0.1 W or 2.2 V) continuous peak, ≤100 V DC blocking capacitor }}
{{Spec | RF Attenuator Range | 0 dB to 50 dB, 2 dB steps }}
{{Spec | RF Attenuator | 0 dB to 50 dB, 2 dB steps }}
{{Spec | Sensitivity (at 300 Hz RBW) | -78 dBmV (-127 dBm); -90 dBmV (-139 dBm) with preamp on }}
{{Spec | Sensitivity | (at 300 Hz RBW) -78 dBmV (-127 dBm); -90 dBmV (-139 dBm) with preamp on }}
{{Spec | Reference Level Range | -70 dBm to +20 dBm; -23 dBmV to +66.9 dBmV }}
{{Spec | Reference Level | -70 dBm to +20 dBm; -23 dBmV to +66.9 dBmV }}
{{Spec | Sweep Speed Range | 2 sec/Div to 1 µs/Div in 1, 2, 5 sequence }}
{{Spec | Sweep Speed | 2 sec/Div to 1 µs/Div in 1, 2, 5 sequence }}
{{Spec | Triggering Modes | Free Run, Internal, External, Line, TV Line, and TV Field }}
{{Spec | Triggering Modes | Free Run, Internal, External, Line, TV Line, and TV Field }}
{{Spec | Displayed Average Noise | −66 dBm to −100 dBm }}
{{Spec | Displayed Average Noise | −66 dBm to −100 dBm }}
Line 39: Line 35:
{{Spec | Dynamic Range | ≥100 dBm }}
{{Spec | Dynamic Range | ≥100 dBm }}
{{Spec | Internal Calibrator | 100 MHz marker at +18.8 dBmV (-30 dBm) }}
{{Spec | Internal Calibrator | 100 MHz marker at +18.8 dBmV (-30 dBm) }}
{{Spec | Included Accessories | PC Software; Power Cord; User Manual; Programmer Manual; Software Reference Manual; Front Cover; 75Ω BNC Connector, 75Ω F Connector; Choice of GPIB or RS-232 (Opt. 03 or Opt. 08)}}
{{Spec | Weight | Standard accessories: 11.34 kg (25 lbs); Basic configuration: 10.21 kg (22.5 lbs) }}
{{Spec | Power | 90 − 132 V<sub>AC</sub>, 48 to 440 Hz; 90 – 250 V<sub>AC</sub>, 48 to 63 Hz.  At 115 V<sub>AC</sub>, 60 Hz, 90 W (standard), 105 W max (with options) }}
{{Spec | Power | 90 − 132 V<sub>AC</sub>, 48 to 440 Hz; 90 – 250 V<sub>AC</sub>, 48 to 63 Hz.  At 115 V<sub>AC</sub>, 60 Hz, 90 W (standard), 105 W max (with options) }}
{{Spec | Operating Atmospherics | Temperatures: 0 °C to +50 °C;  Humidity: MIL-T-28800E; Altitude: Up to 4.6 km (15,000 ft) }}
{{Spec|Features|
* 124 kB NVRAM to save up to 108 waveform displays, 36 front panel setups, -defined key programs, and antenna factor tables.
* Single and delta markers – readout of frequency and amplitude
* Digital storage display
* Acquisition modes: positive peak only, positive/negative peak, Max hold, Waterfall, an analog sweep (when digital storage is off)
* Ensemble averaging
* Internal preamp
* Internal Calibrator
* Alternate reference level units (dBm, dBmV, dBV, dBµV, dBµW, and dBµV/m)
* Direct plot/print (Printers via Opt. 03 GPIB, or RS-232-C interface via Opt. 08).
}}
{{EndSpecs}}
{{EndSpecs}}


==Options==
==Options==
<small>
* Opt. 03: GPIB Interface (cannot be combined with Opt. 08)
* Opt. 03: GPIB Interface (cannot be combined with Opt. 08)
* Opt. 05: 100 kHz to 1.8 GHz External Tracking Generator (includes option 15)
* Opt. 05: 100 kHz to 1.8 GHz External Tracking Generator (includes option 15)
Line 53: Line 58:
* Opt. 30: Rackmount
* Opt. 30: Rackmount
* Opt. 33: Travel Line Package (includes Rain cover, Accessory pouch, Gray filter and Carrying strap)
* Opt. 33: Travel Line Package (includes Rain cover, Accessory pouch, Gray filter and Carrying strap)
* Opt. 34: Portable to Rackmount Adapter.
* Opt. 34: Portable to Rackmount Adapter
</small>
 
===Connectors===
'''Front panel:''' RF INPUT [[BNC connector]] (75 Ω, +20 dBm max, 100 V DC blocking capacitor; OPTION (Unknown connector – Service manual only indicates that the front panel position is for a '''Future Option'''. No internal tracking generators were available like with the other 2710, 2711, and 2712).
 
'''Rear panel:''' J101 [[SMA connector]] (for Opt. 15); J102 EXT TRIG [[BNC connector]] (10 kΩ, 50 V (DC + peak AC)); J103 [[D-sub connector|DE-9 female]] Accessory: Pin 1 Video Input, Pin 2 Ground, Pin 3 Video Output, Pin 6 Sweep Gate, Pin 7 Sweep Output, all others unused);
J104 Opt. 03: [[GPIB connector]] or Opt. 08: [[RS-232 interface]] (Digital Communication Port)
 
==Links==
{{Documents|Link=2714|Class=Application Note}}
 
==See Also==
* [[CSS500]] Cable Television System Software for controlling VM700A, TDC-10, 1450A, 2714, 2715, TV1350, Rohde & Schwarz EMFP


==Pictures==
==Pictures==
Line 66: Line 84:
</gallery>
</gallery>


[[Category: Spectrum Analyzers]]
[[Category:Spectrum analyzers]]

Latest revision as of 10:38, 15 May 2024

Tektronix 2714
1.8 GHz Spectrum Analyzer
Tektronix 2714

Produced from 1993 to 2000

Manuals
Manuals – Specifications – Links – Pictures

The Tektronix 2714 is a CATV spectrum analyzer with a frequency range of 9 kHz to 1.8 GHz.

The 2714 Cable TV Spectrum Analyzer supports CATV-specific functions like automatic positioning of visual aural carriers, visual and aural carrier levels and frequencies for selected channels, survey of system visual and aural carrier levels and frequencies, depth of modulation, aural carrier deviation measurement, visual carrier-to-noise measurement, and visual to carrier amplitude and frequency difference measurements.

Key Specifications

Frequency 9 kHz to 1.8 GHz; Preamp limited to 600 MHz
Frequency Span 1 kHz/div to 100 MHz/div in 1, 2, 5 sequence (plus 0 Hz and MAX: 180 MHz/div)
Resolution Bandwidth –6 dB bandwidth selections are 300 Hz, 1 kHz, 3 kHz, 10 kHz, 30 kHz, 100 kHz, 300 kHz, 1 MHz and 5 MHz
RF Input impedance 75 Ω
Maximum Safe Input Power +70 dBmV (0.1 W or 2.2 V) continuous peak, ≤100 V DC blocking capacitor
RF Attenuator 0 dB to 50 dB, 2 dB steps
Sensitivity (at 300 Hz RBW) -78 dBmV (-127 dBm); -90 dBmV (-139 dBm) with preamp on
Reference Level -70 dBm to +20 dBm; -23 dBmV to +66.9 dBmV
Sweep Speed 2 sec/Div to 1 µs/Div in 1, 2, 5 sequence
Triggering Modes Free Run, Internal, External, Line, TV Line, and TV Field
Displayed Average Noise −66 dBm to −100 dBm
Display Dynamic Range 80 dB
Dynamic Range ≥100 dBm
Internal Calibrator 100 MHz marker at +18.8 dBmV (-30 dBm)
Power 90 − 132 VAC, 48 to 440 Hz; 90 – 250 VAC, 48 to 63 Hz. At 115 VAC, 60 Hz, 90 W (standard), 105 W max (with options)
Features
  • 124 kB NVRAM to save up to 108 waveform displays, 36 front panel setups, -defined key programs, and antenna factor tables.
  • Single and delta markers – readout of frequency and amplitude
  • Digital storage display
  • Acquisition modes: positive peak only, positive/negative peak, Max hold, Waterfall, an analog sweep (when digital storage is off)
  • Ensemble averaging
  • Internal preamp
  • Internal Calibrator
  • Alternate reference level units (dBm, dBmV, dBV, dBµV, dBµW, and dBµV/m)
  • Direct plot/print (Printers via Opt. 03 GPIB, or RS-232-C interface via Opt. 08).

Options

  • Opt. 03: GPIB Interface (cannot be combined with Opt. 08)
  • Opt. 05: 100 kHz to 1.8 GHz External Tracking Generator (includes option 15)
  • Opt. 07: Add 2704 Inverter/2705 Battery Pack
  • Opt. 08: RS-232 Port (cannot be combined with Opt. 03)
  • Opt. 15: Tek 1405 TV Sideband Adapter interface installed at J101 as SMA female
  • Opt. 30: Rackmount
  • Opt. 33: Travel Line Package (includes Rain cover, Accessory pouch, Gray filter and Carrying strap)
  • Opt. 34: Portable to Rackmount Adapter

Connectors

Front panel: RF INPUT BNC connector (75 Ω, +20 dBm max, 100 V DC blocking capacitor; OPTION (Unknown connector – Service manual only indicates that the front panel position is for a Future Option. No internal tracking generators were available like with the other 2710, 2711, and 2712).

Rear panel: J101 SMA connector (for Opt. 15); J102 EXT TRIG BNC connector (10 kΩ, 50 V (DC + peak AC)); J103 DE-9 female Accessory: Pin 1 Video Input, Pin 2 Ground, Pin 3 Video Output, Pin 6 Sweep Gate, Pin 7 Sweep Output, all others unused); J104 Opt. 03: GPIB connector or Opt. 08: RS-232 interface (Digital Communication Port)

Links

Application Notes Referencing 2714

Document Title Authors Year Links
2EW-10598-0.pdf Cable TV Measurements Using the 2714/2715 Spectrum Analyzers 1995

See Also

  • CSS500 Cable Television System Software for controlling VM700A, TDC-10, 1450A, 2714, 2715, TV1350, Rohde & Schwarz EMFP

Pictures