EIP 101A: Difference between revisions
Jump to navigation
Jump to search
No edit summary |
No edit summary |
||
(23 intermediate revisions by 2 users not shown) | |||
Line 1: | Line 1: | ||
{{Plugin Sidebar | {{Plugin Sidebar | ||
|manufacturer=EIP | |||
summary=Microwave Spectrum Analyzer plug-in | | |series=500-series scopes | ||
image= | |type=101A | ||
caption=EIP 101A Microwave Spectrum Analyzer plug-in | |summary=Microwave Spectrum Analyzer plug-in | ||
|image=EIP101A_faceplate.jpg | |||
introduced=(?) | | |caption=EIP 101A Microwave Spectrum Analyzer plug-in | ||
discontinued=(?) | | |introduced=(?) | ||
manuals= | |discontinued=(?) | ||
* | |designers= | ||
|manuals= | |||
* [[Media:EIP 101A_201A_Ops_Svc.pdf|EIP 101A Operation and Instruction Manual w/schematics]] | |||
}} | }} | ||
The '''EIP 101A''' is a Spectrum Analyzer plug-in, designed to work in all Tektronix oscilloscope mainframes which accept letter-type plug-ins. Used in the 201A power supply unit, it works together with all analog oscilloscopes. It covers the frequency range of 0.7- | The '''EIP 101A''' is a receiver-type Spectrum Analyzer plug-in, designed to work in all Tektronix oscilloscope mainframes which accept letter-type plug-ins. | ||
Used in the 201A power supply unit, it works together with all analog oscilloscopes. | |||
It covers the frequency range of 0.7 - 15 GHz in one single band and comprises a three-resonator YIG preselector. | |||
{{BeginSpecs}} | |||
{{Spec | Frequency Range | 1 − 12.4 GHz (useable 0.7 − 15.7 GHz)<sup>*</sup>}} | |||
{{ | {{Spec | Dial Frequency Accuracy | 10 MHz ±3%<sup>**</sup>}} | ||
{{Spec | Video Bandwidth | 1 kHz, 100 kHz, 3 MHz }} | |||
{{Spec | Frequency Response | ±2 dB typical, ±4 dB maximum }} | |||
{{Spec | Calibrated Dispersion | 0 − 11.4 GHz (useable 0 - 15.0 GHz) }} | |||
{{Spec | Dispersion Resettability | 10 MHz }} | |||
{{Spec | Dispersion Dial Accuracy | 10 MHz or ±0.5% whichever is greater }} | |||
{{Spec | Preselection | 3 Resonator YIG filter, nominal 3 dB bandwidth 15 MHz, straight thru leakage >60 dB down, tracking spurious typically >45 dB down<sup>***</sup> }} | |||
{{Spec | Resolution | 15 MHz typ. }} | |||
{{Spec | Image Rejection | Generates no image signals }} | |||
{{Spec | Intermod Rejection | Intermod signals are reduced by the preselector selectivity over conventional microwave diode response }} | |||
{{Spec | Sensitivity | −45 dBm min. (1 kHz bandwidth) | }} | |||
{{Spec | Max. Input Power | +20 dBm }} | |||
{{Spec | Limiting Level | 0.7 − 1.0 GHz -20 dBm typ., 1.0 − 1.2 GHz >+5 dBm, 1.2 − 15.7 GHz >+10 dBm (all at 2 dB compression) }} | |||
{{Spec | Log Accuracy | ±1 dB typ. (<1 mW input) }} | |||
{{Spec | Variable Vertical Log Scale (dB/cm) | Variable and calibratable with internal calibrate signal over the range of 2 dB/cm to 20 dB/cm }} | |||
{{Spec | Log Video output | 0.02 Volts/dB typ. }} | |||
{{Spec | Sweep Rate | 1, 10, 30 Hz }} | |||
{{EndSpecs}} | |||
:<sup>*</sup> Usable frequency coverage is limited by the center frequency dial indicator with the unit set on a center frequency of 12 GHz and dispersion set for 15 GHz, most units will display signals as high as 18 GHz. | |||
:<sup>**</sup> If setting is always approached from 700 MHz, if not, add 5 - 15 MHz for hysteresis. | |||
:<sup>***</sup>Tracking spurious is a second pass band response, always located a fixed number of MHz (approximately 200 MHz) above the normal preselector response. | |||
==Pictures== | ==Pictures== | ||
<gallery> | <gallery> | ||
EIP101A_front_ISO.jpg | EIP101A_front_ISO.jpg | ||
EIP101A_back_ISO.jpg | |||
EIP101A_bottom_front_plate.jpg | |||
EIP101A_bottom_view.jpg | |||
EIP101A_DC_block_and_Detector.jpg | |||
EIP101A_gearbox.jpg | |||
EIP101A_left side.jpg | |||
EIP101A_pcbs.jpg | |||
EIP101A_right_side.jpg | |||
EIP101A_top.jpg | |||
Eip 101a 1.jpeg | |||
</gallery> | </gallery> | ||
[[Category:500 series plugins]] | [[Category:500 series plugins]] | ||
[[Category:Spectrum analyzers]] | [[Category:Spectrum analyzers]] |
Latest revision as of 12:25, 9 January 2024
The EIP 101A is a receiver-type Spectrum Analyzer plug-in, designed to work in all Tektronix oscilloscope mainframes which accept letter-type plug-ins. Used in the 201A power supply unit, it works together with all analog oscilloscopes. It covers the frequency range of 0.7 - 15 GHz in one single band and comprises a three-resonator YIG preselector.
Key Specifications
Frequency Range | 1 − 12.4 GHz (useable 0.7 − 15.7 GHz)* |
---|---|
Dial Frequency Accuracy | 10 MHz ±3%** |
Video Bandwidth | 1 kHz, 100 kHz, 3 MHz |
Frequency Response | ±2 dB typical, ±4 dB maximum |
Calibrated Dispersion | 0 − 11.4 GHz (useable 0 - 15.0 GHz) |
Dispersion Resettability | 10 MHz |
Dispersion Dial Accuracy | 10 MHz or ±0.5% whichever is greater |
Preselection | 3 Resonator YIG filter, nominal 3 dB bandwidth 15 MHz, straight thru leakage >60 dB down, tracking spurious typically >45 dB down*** |
Resolution | 15 MHz typ. |
Image Rejection | Generates no image signals |
Intermod Rejection | Intermod signals are reduced by the preselector selectivity over conventional microwave diode response |
Sensitivity | −45 dBm min. (1 kHz bandwidth) |
Max. Input Power | +20 dBm |
Limiting Level | 0.7 − 1.0 GHz -20 dBm typ., 1.0 − 1.2 GHz >+5 dBm, 1.2 − 15.7 GHz >+10 dBm (all at 2 dB compression) |
Log Accuracy | ±1 dB typ. (<1 mW input) |
Variable Vertical Log Scale (dB/cm) | Variable and calibratable with internal calibrate signal over the range of 2 dB/cm to 20 dB/cm |
Log Video output | 0.02 Volts/dB typ. |
Sweep Rate | 1, 10, 30 Hz |
- * Usable frequency coverage is limited by the center frequency dial indicator with the unit set on a center frequency of 12 GHz and dispersion set for 15 GHz, most units will display signals as high as 18 GHz.
- ** If setting is always approached from 700 MHz, if not, add 5 - 15 MHz for hysteresis.
- ***Tracking spurious is a second pass band response, always located a fixed number of MHz (approximately 200 MHz) above the normal preselector response.