The Tektronix 3S76 is a dual-trace [[sampling oscilloscope|sampling]] plug-in
The Tektronix 3S76 is a dual-trace [[sampling oscilloscope|sampling]] plug-in
for [[560-series scopes]]. The risetime is rated at less than 0.4ns.  Each input
for [[560-series scopes]]. The risetime is rated at less than 0.4ns.  Each input
has a trigger takeoff transformer and a 55ns coaxial delay line.  The trigger source channel is selectable.  The sampling bridge is a four-diode design  
has a trigger takeoff transformer and a 55ns coaxial delay line.  The trigger source channel is selectable.   
using beam-lead GaAs diodes, with a [[7586]] nuvistor buffering its output.   
Each delay line terminates at that channel's sampling bridge,
The sampling pulse is generated by snap-off diode and a 3cm twisted-pair
which is a four-diode design using beam-lead GaAs diodes.
clipping line.
A [[7586]] nuvistor buffers the output of each sampling bridge.   
The sampling pulse is generated by snap-off diode and  
a 3cm twisted-pair clipping line.


* [http://bama.edebris.com/download/tek/3s76/3s76.djvu Tektronix 3S76 Manual (DjVu)]
* [http://bama.edebris.com/download/tek/3s76/3s76.djvu Tektronix 3S76 Manual (DjVu)]
* [http://w140.com/tek_3s76.pdf Tektronix 3S76 Manual (PDF)]
* [http://w140.com/tek_3s76.pdf Tektronix 3S76 Manual (PDF)]

Revision as of 15:05, 18 August 2011

The Tektronix 3S76 is a dual-trace sampling plug-in for 560-series scopes. The risetime is rated at less than 0.4ns. Each input has a trigger takeoff transformer and a 55ns coaxial delay line. The trigger source channel is selectable. Each delay line terminates at that channel's sampling bridge, which is a four-diode design using beam-lead GaAs diodes. A 7586 nuvistor buffers the output of each sampling bridge. The sampling pulse is generated by snap-off diode and a 3cm twisted-pair clipping line.