The '''Tektronix S-42''' is a sampling head for 1300 nm optical signals. It has an [[FC/PC connector|FC/PC fiberoptic connector]] | The '''Tektronix S-42''' is a sampling head for 1300 nm optical signals. It has an [[FC/PC connector|FC/PC fiberoptic connector]] |
Revision as of 19:16, 13 November 2019
Template:Plugin Sidebar 2 The Tektronix S-42 is a sampling head for 1300 nm optical signals. It has an FC/PC fiberoptic connector for the input signal. It is essentially an S-4 sampling head with a photodiode at the input.
Specified accuracy is for connections through 8/125 micron single-mode fibre, although a lens in front of the photodiode allows connections using multi-mode fibre as well, at some degradation of response speed.
The manual shows it being used in a 7S12-driven TDR setup with an external OT503 electro-optic converter as the optical signal source.
Key Specifications
Pulse response | 55 ps |
---|---|
Bandwidth | 6.4 GHz (equiv.) |
Spectral response | 1000 nm – 1700 nm |
Deflection | 50 μW/Div to 50 mW/Div @ 1300 nm |
Noise equivalent power | 42 μWRMS max. |
Max. nondestructive input | 10 mW mean, 125 mW peak |
Features |
|
Internals
The S-42 borrows heavily from the S-4 sampling head, re-using the S-4 strobe, sampler and preamp boards. It uses the same six-diode traveling-wave sampling gate, strobed by 125 ps pulses generated using a snap-off diode driven by an avalanche transistor.
The S-4's SMA input connector is still present internally, and driven electrically in some of the calibration procedures, with the photodiode removed.
To accommodate the photo diode, the S-42's case is longer than the S-4's. When inserted, the S-42 therefore protrudes from the base unit by about two centimeters.
Pictures
-
-
-
-
-
-
block diagram
-
schematic
-
S-42 spectral response
-
-
Description in 1989 Catalog
-