Pacific Measurements 1038: Difference between revisions
Jump to navigation
Jump to search
m (Added images for the Pacific Measurements 1038-NS201) |
m (Added gallery syntax for subsections and headings with clarification for the subsections in the Pictures section) |
||
Line 3: | Line 3: | ||
summary=100kHz to 18.5GHz network analyzer | | summary=100kHz to 18.5GHz network analyzer | | ||
image=Pacific Measurements 1038.jpg | | image=Pacific Measurements 1038.jpg | | ||
caption=Pacific Measurements 1038 with Tek D11 display, PM-H13 and x2 PM-V12 | caption=Pacific Measurements 1038 with Tek D11 display, PM-H13 and x2 PM-V12 plug-ins | | ||
introduced=(?) | | introduced=(?) | | ||
discontinued=(?) | | discontinued=(?) | | ||
Line 74: | Line 74: | ||
==Pictures == | ==Pictures == | ||
===1038-D10 with H13 & (bench model) V11, (rack model) x2 V12 Plug-Ins=== | |||
<gallery> | <gallery> | ||
D10_Mainframe_PM1038_System.jpg | 1038-D10 H/V bench model | D10_Mainframe_PM1038_System.jpg | 1038-D10 H/V bench model | ||
D10_Mainframe_rackmount_PM1038_System.jpg | 1038-D10 H/V rack model | D10_Mainframe_rackmount_PM1038_System.jpg | 1038-D10 H/V rack model | ||
D10_Mainframe_rackmount_PM1038_System_rear.jpg | 1038-D10 rack model, rear | D10_Mainframe_rackmount_PM1038_System_rear.jpg | 1038-D10 rack model, rear | ||
</gallery > | |||
===1038-D11 with H13 & x2 V12 Plug-Ins=== | |||
<gallery> | |||
Pacific Measurements 1038.jpg | 1038-D11 H/V bench model | |||
Pacific Measurements 1038 rear.jpg | 1038-D11 bench model, rear | |||
</gallery > | |||
===1038-D14 with H13 & x2 V12 Plug-Ins=== | |||
<gallery> | |||
PacificMeasurements_1038_D14_H13_V12_front.jpg | 1038-D14 H/V bench model | PacificMeasurements_1038_D14_H13_V12_front.jpg | 1038-D14 H/V bench model | ||
PacificMeasurements_1038_D14_H13_V12_rear.jpg | 1038-D14 bench model, rear | PacificMeasurements_1038_D14_H13_V12_rear.jpg | 1038-D14 bench model, rear | ||
</gallery > | |||
===1038-D14A with N10 Plug-In=== | |||
<gallery> | |||
WavetekD14A_PM1038-N10_front.jpg | 1038-D14A N10 rack model | WavetekD14A_PM1038-N10_front.jpg | 1038-D14A N10 rack model | ||
WavetekD14A_PM1038-N10_back.jpg | 1038-D14A rack model, rear | WavetekD14A_PM1038-N10_back.jpg | 1038-D14A rack model, rear | ||
</gallery > | |||
===1038-D14 with NS20 Plug-In=== | |||
<gallery> | |||
1038_NS20_1200x900_cropped.jpg | 1038-D14 NS20 bench model | 1038_NS20_1200x900_cropped.jpg | 1038-D14 NS20 bench model | ||
</gallery > | |||
===1038-NS201=== | |||
<gallery> | |||
Wavetek_1038_NS201_front_600x450.jpg | 1038-NS201 | Wavetek_1038_NS201_front_600x450.jpg | 1038-NS201 | ||
Wavetek_1038_Ns201_600x450.jpg | 1038-NS201 rear | Wavetek_1038_Ns201_600x450.jpg | 1038-NS201 rear |
Revision as of 15:24, 2 May 2019
The Pacific Measurements 1038 is a 1 or 2 channel scalar network analyzer incorporating a Tektronix D10 (P-7 long persistance phosphor), D11 storage (Split Screen Storage) or [D14] 5000-series scopes display unit mainframe with optional rack mount and IEEE bus.
Options were available for:
- Horizontal Plug-Ins H10 (Sweep drive), H11 (Sweep Drive and memory address circuits), H12 (Sweep drive and ratio channel), H13 (Sweep drive, memory address circuits and ratio channel) and K11 (Replaces H11 and provides amplitude sweep).
- Vertical Plug-Ins V10 (Log amp and signal conditioning), V11 (Log amp, signal conditioning and 256 point memory), V12 (Log amp, signal conditioning and 1024 point memory), V13 (Log amp, signal conditioning, recording of two calibration curves and 1024 point memory) and V20 (Linear amp, 1024 point memory and for use in conjunction with Model 1018B power meter).
- N10 Plug-In (2 Channel Network Analyzer).
- NS20 Plug-In (2 Channel Network Analyzer) with optional NS201 (2 to 18.5 GHz RF Source Module).
- C10 Plug-In Calibration Fixture and Extender Kit for direct calibration and check out of the D14/D14A Mainframe. It has extension cables to permit calibration and maintenance of the H/V system plug-ins.
- Power Input Calibrator Unit (Detector Simulator - P/N 15484) will act as a "Standard" detector to furnish accurate voltages to the H/V system when it is desired to calibrate the amplitde correction circuitry.
Pacific Measurements later became part of Wavetek.
Key Specifications
— H/V System — | |
Offset Range | ±99.9 dB, 3 digits resolution, accuracy ±0.02 dB/10 dB ±0.05 dB |
Drift | @ 0 dBm, <0.01 dB/°C (+15° to +45°C ); @ -40 dBm, <0.5 dB/24 Hrs at constant temperature ) |
Frequency Range | 100 kHz to 18.5 GHz ; with single diode (P/N 12685, 12882, 12883, 13161) detector, 100kHz to 1 GHz or (P/N 13160) 2 GHz ; with single diode (P/N 15272) or balanced (P/N 15285) detector, 1 MHz to 26.5 GHz ; Wavetek Pacific Measurements waveguide detector (P/N 15882), 26.5 GHz to 40.0 GHz ; other waveguide manufacturers, up to 110 GHz |
Power | −50 dBm to +10 dBm |
Dynamic Range | 60 dB (Reference Channel: 40 dB) |
Reading | Insertion Loss/Gain and Return Loss, 0.02 dB resolution |
Noise Floor | −70 dB |
Power | Approx. 100 VA @ 50 − 440 Hz; 100, 120, 240 VAC ±10% |
Dimensions | Bench Mount (H x W x D) 381 x 216 x 483 mm (15 x 8.5 x 19 in.); Rack Mount (H x W x D) 178 x 483 x 483 mm (7 x 19 x 19 in.) |
Weight | 18.2 kg (40 lbs) - Vertical Plug-In 2.3 kg (5 lbs), Horizontal Plug-In 2 kg (4.4 lbs) |
— Power Detectors (P/Ns 15272, 13782 and 13783 unless noted) — | |
Input Impedance | 50 Ω nominal; 75 Ω (P/N 12685, 12882, 13161) |
Input Power | max. +23 dBm (200 mV) peak or average |
Return Loss | >20.8 dB, SWR <1.20 from 100kHz to 1 GHz (P/N 12685, 12882, 12883, 13160, 13161); >25.0 dB, SWR <1.12 from 1 MHz to 1 GHz; >18.0 dB, SWR <1.29 from 2 GHz to 12.4 GHz; >14.0 dB, SWR <1.50 up to 26.5 GHz |
Accuracy | 3% to 18.5 GHz and 5% to 26.5 GHz |
Flatness | ±0.5 dB to 18.5 GHz (P/N 15272); ±1.0 dB 18.5 GHz to 26.5 GHz (P/N 15272) ; ±1.0 dB to 18.5 GHz (P/N 13782) ; ±1.0dB to 18.5 GHz (P/N 13783; ±2.0 dB with type N or APC7 connectors to 18 GHz, ±2.0 dB to 26.5 GHz with APC3.5 connectors |
— Power Detector P/N 15882 — | |
Input Power | max. +20 dBm (100 mV) peak or average |
Return Loss | >10 dB, SWR <1.92 from 26.5 to 40 GHz |
Accuracy | 5% from 26.5 to 40 GHz |
Flatness | ±2.0 dB from 26.5 to 40 GHz |
Temperature | 0° to 50°C operating, -50° to +65°C non-operating |
Accuracy | 0.2 dB to 2 dB from 30 MHz to 26.5 GHz, 0.5 dB to 2 dB from 26.5 GHz to 40 GHz |
Cable Length | 1.5 m (5 ft) |
Weight | Approx. 230 g (8 oz) |
Prices
Please add
Links
- RF Analyzers @ Sphere Research
- OE2IGL: "MAR104 - Mobile Amateur Rain Radar" Project with Return Loss Measurement Example
Pictures
1038-D10 with H13 & (bench model) V11, (rack model) x2 V12 Plug-Ins
-
1038-D10 H/V bench model
-
1038-D10 H/V rack model
-
1038-D10 rack model, rear
1038-D11 with H13 & x2 V12 Plug-Ins
-
1038-D11 H/V bench model
-
1038-D11 bench model, rear
1038-D14 with H13 & x2 V12 Plug-Ins
-
1038-D14 H/V bench model
-
1038-D14 bench model, rear
1038-D14A with N10 Plug-In
-
1038-D14A N10 rack model
-
1038-D14A rack model, rear
1038-D14 with NS20 Plug-In
-
1038-D14 NS20 bench model
1038-NS201
-
1038-NS201
-
1038-NS201 rear
-
1038-NS201
-
1038-NS201 rear