Revision as of 05:55, 9 October 2017
The Tektronix 410 Physiological Monitor is a specialized oscilloscope for medical applications.
It is a small, battery-powered instrument using a magnetically deflected CRT, providing ECG, EEG and pulse measurement modes.
The graticule includes a heart rate scale allowing readings in beats/min (at 50 mm/s).
The 067-0549-00 Input Adapter can be used to calibrate the 410.
Key Specifications
Sweep | 25, 50 or 100 mm/s |
---|---|
Bandwidth | 0.1 to 250 Hz (ECG, AUX) or 100 Hz (EEG) |
Deflection | 10 mm/50 μV (EEG), 20 mm/mV (ECG), 2 mm/mV (AUX) |
Input impedance | 2 MΩ (EEG, ECG) / 20 MΩ (AUX) |
CMRR | 150,000 : 1 |
Battery | 10 size C NiCd cells, operating time 8-12 hours, charging 14-16 hours |
CRT | 5" magnetically deflected, P-7 phosphor |
Links
Photos
-
-
-
-
-
-
Catalog image
-
Inerior front view
-
Inerior front view witout CRT assy
-
Inerior top view
-
Inerior bottom view
-
Inerior back view
-
CRT assembly
-
Vertical amplifier
-
Scan generator, horizontal amplifier and audio channel
-
High Voltage supply