3S3: Difference between revisions
No edit summary |
No edit summary |
||
Line 1: | Line 1: | ||
The Tektronix Type 3S3 is a sampling plug-in for 560-series scopes. | The Tektronix Type 3S3 is a [[sampling_oscilloscope|sampling]] plug-in for [[560-series scopes]]. | ||
It requires P6038 probes which contain miniature sampling gates. | It requires P6038 probes which contain miniature sampling gates. | ||
The risetime is rated at 350 picoseconds. | The risetime is rated at 350 picoseconds. |
Revision as of 20:07, 27 March 2011
The Tektronix Type 3S3 is a sampling plug-in for 560-series scopes. It requires P6038 probes which contain miniature sampling gates. The risetime is rated at 350 picoseconds. The sampling gates use four GaAs diodes. The sampling pulse is produced by the Gate Generator module in the plug-in using a blocking oscillator driving a step recovery diode and a 6.5cm clipping line. Balanced pulses are sent up the cable to the sampling gate. The output of the sampling gate returns to the plug-in through the same cable, and enters the preamp. The same blocking oscillator that generates the sampling pulses also generates the strobe for the memory circuit. The sampling efficiency is specified to be 2%. The 3S3 is recommended in the Type 111 manual for checking the output pulse risetime.
-
Front View
-
Left View
-
Right View
-
Top View
-
Cards Removed
-
Gate Generator
-
Gate Generator
-
Memory
-
Memory
-
Inverter
-
Inverter
-
Dual Trace
-
Dual Trace
-
Vertical Amplifier
-
Vertical Amplifier
-
Pulse testing an amplifier
-
Simplified block diagram
-
Simplified input circuitry
-
Sampling pulse generator