Line 12: Line 12:
* [[Media:Pacific_Measurements_PM1038_Swept_Power_Meter_Instruction_Manual-D14.pdf|PMI 1038-D14 Mainframe Instruction Manual (PDF)]]
* [[Media:Pacific_Measurements_PM1038_Swept_Power_Meter_Instruction_Manual-D14.pdf|PMI 1038-D14 Mainframe Instruction Manual (PDF)]]
* [[Media:1038_H13_Operation_Maintenance_Manual_full.pdf |Wavetek PMI 1038-H13 Operating and Maintenance Manual (PDF)]]
* [[Media:1038_H13_Operation_Maintenance_Manual_full.pdf |Wavetek PMI 1038-H13 Operating and Maintenance Manual (PDF)]]
* [[Media:1038_V10_V11_V12_Instruction_Manual_full.pdf |PMI 1038-V10, -V11 and -V12 Instruction Manual (PDF)]]
* [[Media:Pacific_Measurements_PM1038_Swept_Power_Meter_App_Note18.pdf|PMI 1038 Swept Measurement System Application Note 18 (PDF)]]
* [[Media:Pacific_Measurements_PM1038_Swept_Power_Meter_App_Note18.pdf|PMI 1038 Swept Measurement System Application Note 18 (PDF)]]
* [[Media:Pacific_Measurements_PM1038_Swept_Power_Meter_App_Note19.pdf|PMI 1038 Swept Measurement System Application Note 19 (PDF)]]
* [[Media:Pacific_Measurements_PM1038_Swept_Power_Meter_App_Note19.pdf|PMI 1038 Swept Measurement System Application Note 19 (PDF)]]
Line 44: Line 45:
{{Spec | Noise Floor | −70 dB }}
{{Spec | Noise Floor | −70 dB }}
{{Spec | Power | Approx. 100 VA @ 50 − 440 Hz; 100, 120, 240 V<sub>AC</sub> ±10% }}
{{Spec | Power | Approx. 100 VA @ 50 − 440 Hz; 100, 120, 240 V<sub>AC</sub> ±10% }}
{{Spec | Dimensions | Bench Mount (H x W x D) | 381 x 216 x 483 mm (15 x 8.5 x 19 in.);<br />Rack Mount (H x W x D) | 178 x 483 x 483 mm (7 x 19 x 19 in.); }}
{{Spec | Dimensions | Bench Mount (H x W x D) 381 x 216 x 483 mm (15 x 8.5 x 19 in.);<br />Rack Mount (H x W x D) 178 x 483 x 483 mm (7 x 19 x 19 in.) }}
{{Spec | Weight | 18.2 kg (40 lbs) - Vertical Plug-In 2.3 kg (5 lbs), Horizontal Plug-In | 2 kg (4.4 lbs) }}
{{Spec | Weight | 18.2 kg (40 lbs) - Vertical Plug-In 2.3 kg (5 lbs), Horizontal Plug-In 2 kg (4.4 lbs) }}
{{SpecGroup | Power Detectors (P/Ns 15272, 13782 and 13783 unless noted) }}
{{SpecGroup | Power Detectors (P/Ns 15272, 13782 and 13783 unless noted) }}
{{Spec | Input Impedance | 50 Ω nominal }}
{{Spec | Input Impedance | 50 Ω nominal }}

Revision as of 09:48, 30 April 2019

Manuals – Specifications – Links – Pictures

The Pacific Measurements 1038 is a 1 or 2 channel scalar network analyzer incorporating a Tektronix D10 (P-7 long persistance phosphor), D11 storage (Split Screen Storage) or [D14] 5000-series scopes display unit mainframe with optional rack mount and IEEE bus.

Options were available for:

  • Horizontal Plug-Ins H10 (Sweep drive), H11 (Sweep Drive and memory address circuits), H12 (Sweep drive and ratio channel), H13 (Sweep drive, memory address circuits and ratio channel) and K11 (Replaces H11 and provides amplitude sweep).
  • Vertical Plug-Ins V10 (Log amp and signal conditioning), V11 (Log amp, signal conditioning and 256 point memory), V12 (Log amp, signal conditioning and 1024 point memory), V13 (Log amp, signal conditioning, recording of two calibration curves and 1024 point memory) and V20 (Linear amp, 1024 point memory and for use in conjunction with Model 1018B power meter).
  • N10 Plug-In (2 Channel Network Analyzer).
  • NS20 Plug-In (2 Channel Network Analyzer) with optional NS201 (2 to 18.5 GHz RF Source Module).
  • C10 Plug-In Calibration Fixture and Extender Kit for direct calibration and check out of the D14/D14A Mainframe. It has extension cables to permit calibration and maintenance of the H/V system plug-ins.
  • Power Input Calibrator Unit (Detector Simulator - P/N 15484) will act as a "Standard" detector to furnish accurate voltages to the H/V system when it is desired to calibrate the amplitde correction circuitry.

Pacific Measurements later became part of Wavetek.

Key Specifications

     — H/V System
Offset Range ±99.9 dB, 3 digits resolution, accuracy ±0.02 dB/10 dB ±0.05 dB
Drift @ 0 dBm, <0.01 dB/°C (+15° to +45°C );
@ -40 dBm, <0.5 dB/24 Hrs at constant temperature )
Frequency Range 100 kHz to 18.5 GHz ;
with single diode (P/N 15272) or balanced (P/N 15285) detector, 1 MHz to 26.5 GHz ;
Wavetek Pacific Measurements waveguide detector (P/N 15882), 26.5 GHz to 40.0 GHz ;
other waveguide manufacturers, up to 110 GHz
Power −50 dBm to +10 dBm
Dynamic Range 60 dB (Reference Channel: 40 dB)
Reading Insertion Loss/Gain and Return Loss, 0.02 dB resolution
Noise Floor −70 dB
Power Approx. 100 VA @ 50 − 440 Hz; 100, 120, 240 VAC ±10%
Dimensions Bench Mount (H x W x D) 381 x 216 x 483 mm (15 x 8.5 x 19 in.);
Rack Mount (H x W x D) 178 x 483 x 483 mm (7 x 19 x 19 in.)
Weight 18.2 kg (40 lbs) - Vertical Plug-In 2.3 kg (5 lbs), Horizontal Plug-In 2 kg (4.4 lbs)
     — Power Detectors (P/Ns 15272, 13782 and 13783 unless noted)
Input Impedance 50 Ω nominal
Input Power max. +23 dBm (200 mV) peak or average
Return Loss >25.0 dB, SWR <1.12 from 1 MHz to 1 GHz;
>18.0 dB, SWR <1.29 from 2 GHz to 12.4 GHz;
>14.0 dB, SWR <1.50 up to 26.5 GHz
Accuracy 3% to 18.5 GHz and 5% to 26.5 GHz
Flatness ±0.5 dB to 18.5 GHz (P/N 15272);
±1.0 dB 18.5 GHz to 26.5 GHz (P/N 15272) ;
±1.0 dB to 18.5 GHz (P/N 13782) ;
±1.0dB to 18.5 GHz (P/N 13783;
±2.0 dB with type N or APC7 connectors to 18 GHz, ±2.0 dB to 26.5 GHz with APC3.5 connectors
     — Power Detector P/N 15882
Input Power max. +20 dBm (100 mV) peak or average
Return Loss >10 dB, SWR <1.92 from 26.5 to 40 GHz
Accuracy 5% from 26.5 to 40 GHz
Flatness ±2.0 dB from 26.5 to 40 GHz
Temperature 0° to 50°C operating, -50° to +65°C non-operating
Accuracy 0.2 dB to 2 dB from 30 MHz to 26.5 GHz, 0.5 dB to 2 dB from 26.5 GHz to 40 GHz
Cable Length 1.5 m (5 ft)
Weight Approx. 230 g (8 oz)

Prices

Please add

Links