Pacific Measurements 1038: Difference between revisions
(Removed the pipe "|" and open and closed brackets "[ ]" symbols due to syntax not required when using URL link for the Tektronix D10 Instruction Manual) |
(Added newer version Wavetek Microwave 1038-V13 Operating and Maintenance Manual as well as corporate naming reference.) |
||
Line 17: | Line 17: | ||
* [[Media:1038_V10_V11_V12_Instruction_Manual_full.pdf |PMI 1038-V10, -V11 and -V12 Instruction Manual (PDF)]] | * [[Media:1038_V10_V11_V12_Instruction_Manual_full.pdf |PMI 1038-V10, -V11 and -V12 Instruction Manual (PDF)]] | ||
* [[Media:PMI_1038-V13_Operating_and_Maintenance_Manual.pdf |PMI 1038-V13 Operating and Maintenance Manual (PDF)]] | * [[Media:PMI_1038-V13_Operating_and_Maintenance_Manual.pdf |PMI 1038-V13 Operating and Maintenance Manual (PDF)]] | ||
* [[Media:Wavetek_1038_V13_Operating_and_Maintenance_Manual.pdf |Wavetek Microwave 1038-V13 Operating and Maintenance Manual (PDF)]] | |||
* [[Media:Pacific_Measurements_PM1038_Swept_Power_Meter_App_Note18.pdf|PMI 1038 Swept Measurement System Application Note 18 (PDF)]] | * [[Media:Pacific_Measurements_PM1038_Swept_Power_Meter_App_Note18.pdf|PMI 1038 Swept Measurement System Application Note 18 (PDF)]] | ||
* [[Media:Pacific_Measurements_PM1038_Swept_Power_Meter_App_Note19.pdf|PMI 1038 Swept Measurement System Application Note 19 (PDF)]] | * [[Media:Pacific_Measurements_PM1038_Swept_Power_Meter_App_Note19.pdf|PMI 1038 Swept Measurement System Application Note 19 (PDF)]] | ||
Line 38: | Line 39: | ||
* Power input calibrator unit (detector simulator - P/N 15484) will act as a "standard" detector to furnish accurate voltages to the H/V system when it is desired to calibrate the amplitude correction circuitry | * Power input calibrator unit (detector simulator - P/N 15484) will act as a "standard" detector to furnish accurate voltages to the H/V system when it is desired to calibrate the amplitude correction circuitry | ||
Pacific Measurements later became part of [[Wavetek]]. | Pacific Measurements later became part of [[Wavetek]] as Wavetek Pacific Measurements and later as Wavetek Microwave. | ||
{{BeginSpecs}} | {{BeginSpecs}} |
Revision as of 14:06, 13 May 2019
The Pacific Measurements 1038 is a 1 or 2 channel scalar network analyzer, some models of which incorporate a Tektronix 5000-series display unit − a D10 long-persistance display (P-7 phosphor) or a D11 split-screen storage display.
The PM1038-D14 has a display unit with digital storage. From manual photos and diagrams, the D14 would appear not to be a Tek OEM unit (see also Introduction to the 5000-Series Oscilloscopes regarding 5114/D14).
The PM1038 had an optional GPIB interface and was available in benchtop or rackmount configurations.
Options:
- Horizontal plug-ins: H10 (sweep drive), H11 (sweep Drive and memory address circuits), H12 (sweep drive and ratio channel), H13 (sweep drive, memory address circuits and ratio channel) and K10/K11 (replaces H11 and provides amplitude sweep).
- Vertical plug-ins: V10 (log amp and signal conditioning), V11 (log amp, signal conditioning and 256 point memory), V12 (log amp, signal conditioning and 1024 point memory), V13 (log amp, signal conditioning, recording of two calibration curves and 1024 point memory) and V20 (linear amp, 1024 point memory and for use in conjunction with Model 1018B power meter).
- N10 plug-in (dual channel network analyzer)
- NS20 plug-in (dual channel network analyzer) with optional NS201 (2 to 18.5 GHz RF source)
- C10 plug-in calibration fixture and extender kit for direct calibration and check-out of the D14/D14A Mainframe. It has extension cables to permit calibration and maintenance of the H/V system plug-ins
- Power input calibrator unit (detector simulator - P/N 15484) will act as a "standard" detector to furnish accurate voltages to the H/V system when it is desired to calibrate the amplitude correction circuitry
Pacific Measurements later became part of Wavetek as Wavetek Pacific Measurements and later as Wavetek Microwave.
Key Specifications
— H/V System — | |
Offset Range | ±99.9 dB, 3 digits resolution, accuracy ±0.02 dB/10 dB ±0.05 dB |
Drift | @ 0 dBm, <0.01 dB/°C (+15° to +45°C ); @ −40 dBm, <0.5 dB/24 Hrs at constant temperature ) |
Frequency Range | 100 kHz to 18.5 GHz ; with single diode (P/N 12685, 12882, 12883, 13161) detector, 100 kHz to 1 GHz or (P/N 13160) 2 GHz ; with single diode (P/N 15272) or balanced (P/N 15285) detector, 1 MHz to 26.5 GHz ; Wavetek Pacific Measurements waveguide detector (P/N 15882), 26.5 GHz to 40.0 GHz ; other waveguide manufacturers, up to 110 GHz |
Power | −50 dBm to +10 dBm |
Dynamic Range | 60 dB (Reference Channel: 40 dB) |
Reading | Insertion Loss/Gain and Return Loss, 0.02 dB resolution |
Noise Floor | −70 dB |
Power | Approx. 100 VA @ 50 − 440 Hz; 100, 120, 240 VAC ±10% |
Dimensions | benchtop (H x W x D) 381 x 216 x 483 mm (15 x 8.5 x 19 in.); rackmount (H x W x D) 178 x 483 x 483 mm (7 x 19 x 19 in.) |
Weight | 18.2 kg (40 lbs) - vertical plug-in 2.3 kg (5 lbs), horizontal plug-in 2 kg (4.4 lbs) |
— Power Detectors (P/Ns 15272, 13782 and 13783 unless noted) — | |
Input Impedance | 50 Ω nominal; 75 Ω (P/N 12685, 12882, 13161) |
Input Power | max. +23 dBm (200 mV) peak or average |
Return Loss | >20.8 dB, SWR <1.20 from 100 kHz to 1 GHz (P/N 12685, 12882, 12883, 13160, 13161); >25.0 dB, SWR <1.12 from 1 MHz to 1 GHz; >18.0 dB, SWR <1.29 from 2 GHz to 12.4 GHz; >14.0 dB, SWR <1.50 up to 26.5 GHz |
Accuracy | 3% to 18.5 GHz and 5% to 26.5 GHz |
Flatness | ±0.5 dB to 18.5 GHz (P/N 15272); ±1.0 dB 18.5 GHz to 26.5 GHz (P/N 15272) ; ±1.0 dB to 18.5 GHz (P/N 13782) ; ±1.0 dB to 18.5 GHz (P/N 13783; ±2.0 dB with type N or APC7 connectors to 18 GHz, ±2.0 dB to 26.5 GHz with 3.5 mm connectors |
— Power Detector P/N 15882 — | |
Input Power | max. +20 dBm (100 mV) peak or average |
Return Loss | >10 dB, SWR <1.92 from 26.5 to 40 GHz |
Accuracy | 5% from 26.5 to 40 GHz |
Flatness | ±2.0 dB from 26.5 to 40 GHz |
Temperature | 0° to 50°C operating, −50° to +65°C non-operating |
Accuracy | 0.2 dB to 2 dB from 30 MHz to 26.5 GHz, 0.5 dB to 2 dB from 26.5 GHz to 40 GHz |
Cable Length | 1.5 m (5 ft) |
Weight | Approx. 230 g (8 oz) |
Prices
Please add
See Also
5000 Series plug-in interface
5103
5110 (5103/D10)
5111 (5103/D11)
Links
- RF Analyzers @ Sphere Research
- OE2IGL: "MAR104 - Mobile Amateur Rain Radar" Project with Return Loss Measurement Example
Pictures
1038-D10
-
1038-D10 H/V bench model with H13 and V11 plug-ins
-
1038-D10 H/V rack model with H13 and two V12 plug-ins
-
1038-D10 rack model, rear
1038-D11
-
1038-D11 H/V bench model with H13 and two V12 plug-ins
-
1038-D11 bench model, rear
1038-D14
-
1038-D14 H/V bench model with H13 and two V12 plug-ins
-
1038-D14 bench model, rear
1038-D14A
-
1038-D14A N10 rack model with N10 plug-in
-
1038-D14A rack model, rear
1038-D14
-
1038-D14 bench model with NS20 plug-in
1038-NS201
-
1038-NS201
-
1038-NS201 rear
-
1038-NS201
-
1038-NS201 rear