{{Spec | Maximum Voltage | ±25 V (DC + pk AC) between signal and common of the same channel}} | |||
{{Spec | Attenuation | 1:1 or 10:1 }} | |||
{{Spec | Differential Signal Range | ±850 mV (1:1), ±8 V (10:1)}} | |||
{{Spec | Common-Mode Signal Range |±7 V}} | |||
{{Spec | Common-Mode Rejection Ratio | 60 dB @ 1MHz, 30 dB @ 1GHz (1:1), 45 dB @ 1MHz, 18 dB @ 1GHz (10:1)}} | |||
{{Spec | Differential Offset Range | ±0.7V (1:1), ±7V (10:1)}} | |||
{{Spec | Differential Input Resistance | 200 kΩ}} | |||
{{Spec | Differential Input Capacitance | <1 pF}} | |||
{{Spec | Delay | 20 ns }} | |||
{{Spec | Output | 50 Ω output drives 1 MΩ // 15-20 pF scope input; scope bandwidth should be ≥200 MHz }} | |||
{{EndSpecs}} | |||
==Manual== | ==Manual== |
Revision as of 17:18, 19 September 2020
The Tektronix P6246, P6247 and P6248 are active differential probes with a Tekprobe BNC connector. Bandwidth is 1.5 GHz for the 6248, 1 GHz for the 6247 and 400 MHz for the 6246.
Key Specifications
Maximum Voltage | ±25 V (DC + pk AC) between signal and common of the same channel |
---|---|
Attenuation | 1:1 or 10:1 |
Differential Signal Range | ±850 mV (1:1), ±8 V (10:1) |
Common-Mode Signal Range | ±7 V |
Common-Mode Rejection Ratio | 60 dB @ 1MHz, 30 dB @ 1GHz (1:1), 45 dB @ 1MHz, 18 dB @ 1GHz (10:1) |
Differential Offset Range | ±0.7V (1:1), ±7V (10:1) |
Differential Input Resistance | 200 kΩ |
Differential Input Capacitance | <1 pF |
Delay | 20 ns |
Output | 50 Ω output drives 1 MΩ // 15-20 pF scope input; scope bandwidth should be ≥200 MHz |