TDS3014: Difference between revisions
No edit summary |
mNo edit summary |
||
Line 16: | Line 16: | ||
Besides the channel inputs and the EXT TRIG input (two-channel models only), a [[Floppy]] Disk Drive and Centronics [[Parallel printer interface]] was available on the original and B version of this oscilloscope series until the C version which replaced the [[Floppy]] with a [[USB connector]] drive, and [[Parallel printer interface]] was removed altogether. Optional Communication Modules were available such as the TDS3GM (GPIB/RS-232), TDS3VM ([[VGA Connector]]/[[RS-232 interface]]), TDS3EM ([[Ethernet]]/[[RS-232 interface]]), TDS3GV ([[GPIB]], [[RS-232 interface]]/[[VGA Connector]]), and the TDS3SDI 601 Serial/Digital Video Module. | Besides the channel inputs and the EXT TRIG input (two-channel models only), a [[Floppy]] Disk Drive and Centronics [[Parallel printer interface]] was available on the original and B version of this oscilloscope series until the C version which replaced the [[Floppy]] with a [[USB connector]] drive, and [[Parallel printer interface]] was removed altogether. Optional Communication Modules were available such as the TDS3GM (GPIB/RS-232), TDS3VM ([[VGA Connector]]/[[RS-232 interface]]), TDS3EM ([[Ethernet]]/[[RS-232 interface]]), TDS3GV ([[GPIB]], [[RS-232 interface]]/[[VGA Connector]]), and the TDS3SDI 601 Serial/Digital Video Module. | ||
Up to four Applications modules can be added into the small door in the upper right corner to transform the oscilloscope into a specialized instrument for limit testing, telecommunications mask testing, and video troubleshooting. The TDS3AAM Advanced Analysis Module, TDS3LIM Limit Testing Module, TDS3TMT Telecommunications Mask Testing Module, and the TDS3VID Extended Video Analysis Module were available for added functionality. The TDS3TRG (Advanced Trigger) and TDS3FFT (Fast Fourier Transform) application modules were available in the original model of this oscilloscope series but made standard on the B and higher versions. | Up to four Applications modules can be added into the small door in the upper right corner to transform the oscilloscope into a specialized instrument for limit testing, telecommunications mask testing, and video troubleshooting. The [[TDS3AAM]] Advanced Analysis Module, TDS3LIM Limit Testing Module, TDS3TMT Telecommunications Mask Testing Module, and the TDS3VID Extended Video Analysis Module were available for added functionality. The TDS3TRG (Advanced Trigger) and TDS3FFT (Fast Fourier Transform) application modules were available in the original model of this oscilloscope series but made standard on the B and higher versions. | ||
There are many options for signal acquisition, measurement, and analysis. The Getting Started section of the User Manual is a great place to become familiar with the many functions available in the use of this sophisticated instrument. | There are many options for signal acquisition, measurement, and analysis. The Getting Started section of the User Manual is a great place to become familiar with the many functions available in the use of this sophisticated instrument. |
Revision as of 17:00, 31 December 2020
The Tektronix TDS3014 is a four-channel, 100 MHz, 1.25 GS/s digital phosphor oscilloscope with a TFT color screen. This series of scopes combines 3,600 wfms/s continuous waveform capture rate and real-time intensity grading so you can see the problem and solve it. This series of scope can run without AC power for up to three hours when TDS3BATC rechargeable lithium ion battery pack is installed.
Besides the channel inputs and the EXT TRIG input (two-channel models only), a Floppy Disk Drive and Centronics Parallel printer interface was available on the original and B version of this oscilloscope series until the C version which replaced the Floppy with a USB connector drive, and Parallel printer interface was removed altogether. Optional Communication Modules were available such as the TDS3GM (GPIB/RS-232), TDS3VM (VGA Connector/RS-232 interface), TDS3EM (Ethernet/RS-232 interface), TDS3GV (GPIB, RS-232 interface/VGA Connector), and the TDS3SDI 601 Serial/Digital Video Module.
Up to four Applications modules can be added into the small door in the upper right corner to transform the oscilloscope into a specialized instrument for limit testing, telecommunications mask testing, and video troubleshooting. The TDS3AAM Advanced Analysis Module, TDS3LIM Limit Testing Module, TDS3TMT Telecommunications Mask Testing Module, and the TDS3VID Extended Video Analysis Module were available for added functionality. The TDS3TRG (Advanced Trigger) and TDS3FFT (Fast Fourier Transform) application modules were available in the original model of this oscilloscope series but made standard on the B and higher versions.
There are many options for signal acquisition, measurement, and analysis. The Getting Started section of the User Manual is a great place to become familiar with the many functions available in the use of this sophisticated instrument.
Key Specifications
Bandwidth and Channels | 100 MHz; Analog Bandwidth Limit is 20 MHz; 4 Channels |
---|---|
Vertical Sensitivity | 1 mV to 10 V/div (1 MΩ), 1 mV to 1 V/div (50 Ω) |
Time Base Range | 4 ns to 10 s/div |
Sample Rate | 1.25 GS/s on each channel |
Record Length | 10 K points |
Waveform Measurements | 25 Automatic, waveform add, subtract, divide, and multiply math functions, Fast Fourier Transform (FFT), Advanced Math (Integrate, Differentiate, Define extensive algebraic expressions, math functions, scalars, user adjustable variables) |