==R10 Repair== | ==R10 Repair== | ||
Fault pattern: | Fault pattern: V<sub>DC</sub>, V<sub>AC</sub> and resistance (exept the lowest range) are 2% to low. | ||
Fault: R10. Some resistors are to low (900 kΩ and 90 kΩ). The "stuff" inside has become liquid. The liquid has a low resistance. | Fault: R10. Some resistors are to low (900 kΩ and 90 kΩ). The "stuff" inside has become liquid. The liquid has a low resistance. |
Revision as of 10:05, 2 September 2023
Switch Repair
The contact bar in switches can break. Replacing the contact bar and contacts can be done without desoldering the switches. Refer to the the pictures and the repair section of switch assembly 260-1971-00.
R10 Repair
Fault pattern: VDC, VAC and resistance (exept the lowest range) are 2% to low.
Fault: R10. Some resistors are to low (900 kΩ and 90 kΩ). The "stuff" inside has become liquid. The liquid has a low resistance.
Solution: Open R10. Clean R10 inside with isopropanol and in a ultrasonic bath.
See Also
Pictures
-
Replace the push button plunger and spring contacts on a DM501A
-
Defect push button plunger on top and new one below