067-0587-01 Staircase accuracy

± 0.3% with no error in mainframe ±50 V supply

067-0587-02 Amplitude

2 to 10 Div

067-0587-02 Flatness

± 3% from 3 MHz to 1 GHz

067-0587-02 Input impedance

50 Ω

067-0587-02 Input range

0.5 to 1 Vp-p

067-0587-02 Position

Top at least ±5 Div from center of screen

067-0587-02 Pretrig Out

0.5 V into 50 Ω, 55 to 95 ns lead time (internally variable)

067-0587-02 Repetition rate

10 Hz to 1 MHz in decade steps, 0.1 %

067-0587-02 Rise time

< 150 ps

067-0587-02 Staircase accuracy

± 0.3% in ± 3 Div, ± 0.5% in ± 4 Div, Linearity ± 0.2% of 10 Div

067-0587-10 Amplitude

2 to 10 Div

067-0587-10 Flatness

± 3% from 3 MHz to 1 GHz

067-0587-10 Input impedance

50 Ω

067-0587-10 Input range

0.5 to 1 Vp-p

067-0587-10 Position

Top at least ±5 Div from center of screen

067-0587-10 Pretrig Out

0.5 V into 50 Ω, 55 to 95 ns lead time (internally variable)

067-0587-10 Repetition rate

10 Hz to 1 MHz in decade steps, 0.1%

067-0587-10 Rise time

< 150 ps

067-0587-10 Staircase accuracy

±0.3% in ±3 Div, ±0.5% in ±4 Div, Linearity ±0.2% of 10 Div

067-0625-00 Output

Into floating 2 MΩ load or differential oscilloscope with 1 MΩ inputs

067-0625-00 Sensitivity

10 ±1 mV/% input change at 1.2 Vp-p

067-0625-00 Temperature Range

23°C ±10°C

067-0650-00 Amplitude

0.5 Vp-p to 4 Vp-p into 50 Ω in 0.5 V increments;
Continuously variable between calibrated steps; Max. 4.2 Vp-p

067-0650-00 Amplitude accuracy

9 MHz: Within 2% of indicated amplitude; Constant amplitude 220 MHz to 900 MHz: Within 5% of 9 MHz amplitude

067-0650-00 Frequency range

220 MHz to 900 MHz, or 9 MHz

067-0650-00 Frequency accuracy

Within 2% of indicated frequency; 9 MHz within 5%

067-0650-00 Output impedance

50 Ω ±1%

067-0650-00 Temperature

0 °C to 50 °C

067-0938-00 Notch depth

≥100 dB at 10 Hz and 20 Hz, ≥60 dB otherwise

067-0938-00 Notch frequency

10 Hz, 20 Hz, 100 Hz, 1 kHz, 10 kHz, 20 kHz, 50 kHz, 100 kHz; ±2%

067-1232-00 Attenuation

Input to REF port, 7.8±0.5 dB; REF to 75 Ω port, 3.5±0.2 dB

067-1232-00 Frequency range

DC to 1 GHz

067-1338-00 Output

positive-going step of ~250 mV

067-1338-00 Rise time

max. 19.5 ps, actual value specified on a per-unit basis

067-1429-00 Bandwidth

>10MHz

067-1429-00 Gain

×10 ±2% ( ×1 using passive ×10 probe)

067-1429-00 Input impedance

1 MΩ //20 pF

067-1429-00 Output impedance

75 Ω

107 Output

rectangular waveform, 100−500 mVp-p (variable) into 50 Ω

107 Power

100 W

107 Repetition rate

400 kHz to 1 MHz (variable, uncalibrated)

107 Rise time

less than 3 ns (fixed) when terminated into 50 Ω

107 Weight

5.9 kg (13 pounds)

109 Pulse amplitude

Internal supply: 0 to 50 V variable in three ranges (0.5 V, 5 V, 50 V), positive or negative, or external up to ±300 V

109 Pulse frequency

550 to 720 Hz

109 Pulse length

0.5 ns to 100 ns with a single charge line between the two charge line connectors, or up to 300 ns at half pulse rate with one open-ended charge line (other connector grounded)

109 Rise time

250 ps

10A1 Bandwidth

DC to 45 MHz (35 MHz in maximum sensitivity)

10A1 CMRR

20,000:1 below 100 kHz

10A1 Comparison voltage

−6 V to +6 V ±(0.1% + 5 mV) 0 °C to 40 °C