Key Specifications

Collector Supply

Sweep Modes Normal mode: AC (at line frequency); positive-or negative-going full wave rectified AC.

DC mode: positive or negative DC.

DC Mode Ripple No-load: 2% or less of voltage, or 0.1% or less of full range voltage.
Voltages Accuracy Peak open circuit voltages on all ranges within +35% and -5%.
Voltage Range 15 V 75 V 350 V 1500 V
Maximum Peak Current (Normal Mode)¹ 10 A 2 A 0.5 A 0.1 A
Peak Current (Step Generator in Pulsed Steps Mode) At least 20 A At least 4 A At least 1 A At least 0.2 A
Minimum Series Resistance 0.3 Ω 6.5 Ω 140 Ω 3 kΩ
Maximum Series Resistance 65 kΩ 1.4 MΩ 6.5 MΩ 6.5 MΩ
Series Resistances Available 0.3 Ω, 1.4 Ω, 6.5 Ω, 30 Ω, 140 Ω, 650 Ω, 3 kΩ, 14 kΩ, 65 kΩ, 300 kΩ, 1.4 MΩ, and 6.5 MΩ, all within 5% or 0.1 Ω.
Peak Power Watts Settings 0.1 W, 0.5 W, 2.2 W, l0 W, 50 W and 220 W. Derived from nominal peak open circuit collector voltages and nominal series resistance values at nominal line voltage.
Safety Interlock When MAX PEAK VOLTS switch is set to either 75, 350 or 1500, a protective box must be in place over test terminals and its lid closed before voltage can be applied. Amber light on indicates interlock is open & Red light on indicates voltage is being applied to test terminals.
Looping Compensation Cancels stray capacitance between collector test terminal and ground in Standard Test Fixture and all Standard Test Fixture Accessories.

Step Generator

Accuracy (Current or Voltage Steps, lncluding Offset)
Incremental Accuracy Within 5% between any two steps, without .1 X STEP MULT button pressed; within 10% with .1 X STEP MUL T button pressed.
Absolute Accuracy Within 2% of total output, including any amount of offset, or 1% of AMPLITUDE switch setting, whichever is greater.
Step (Current or Voltage) Amplitudes One times or 0.1 times (with .1 X STEP MUL T button pressed) the AMPLITUDE switch setting.
OFFSET MULT Control Range Continuously variable from 0 to 10 times AMPLITUDE switch setting, either aiding or opposing the step generator polarity.
Current Mode
AMPLITUDE Switch Range 200 mA to 50 nA, in 1-2-5 sequence.
Maximum Current (Steps and Aiding Offset)² 20 times AMPLITUDE switch setting, except 10 times switch setting when switch is set to 200 mA, and 15 times switch setting when the switch is set to 100 mA.
Maximum Voltage (Steps and Aiding Offset) At least 10 V.
Maximum Opposing Offset Current Whichever is less: 10 times AMPLITUDE switch setting, or between 10 mA and 20 mA.
Maximum Opposing Voltage Between 1 V and 3 V.
Ripple Plus Noise 0.5% or less of AMPLITUDE switch setting or 1 nA, peak to peak.
Voltage Mode
AMPLITUDE Switch Range 50 mV to 2 V, in 1-2-5 sequence.
Maximum Voltage (Steps and Aiding Offset) 20 times AMPLITUDE switch setting.
Maximum Current (Steps and Aiding Offset) At least 2 A at 10 V or less, decreasing linearly to 10 mA at 40 V.
Short Circuit Current Limiting (Steps and Aiding Offset) 20 mA, 100 mA, 500 mA, +100%-0%; 2 A +50%-0%; as selected by CURRENT LIMIT switch.
Maximum Opposing Offset Voltage 10 times AMPLITUDE switch setting.
Maximum Opposing Current Limited at between 5 mA and 20 mA
Ripple Plus Noise 0.5% or less of AMPLITUDE switch setting, or 2 mV, peak to peak.
Step Rates (Front panel RATE button labels in parentheses.) 1 times (.5X), 2 times (NORM) and 4 times (2X) line frequency. Steps occur at zero collector voltage when .5X or NORM RATE buttons are pressed. and also at peak voltage when 2X RATE button is pressed. Steps occur at collector voltage peak and at normal rate when .5X and 2X RATE buttons are pressed together.
Pulsed Steps Pulsed steps 80 μS wide within +20%, -5% or 300 μS wide within +5%, -15% produced whenever one of the PULSED STEPS buttons is pressed. Pulsed steps can be produced only at normal and .5 times normal rates. Collector Supply mode automatically becomes DC when either the 300 μS or 80 μS PULSED STEPS button is pressed unless POLARITY switch is set to AC. If the 300 μS and 80 μS PULSED STEPS buttons are pressed together, 300 μS pulsed steps are produced, but collector supply mode does not change.
Steps and Offset Polarity Corresponds with collector supply polarity (positive going when POLARITY switch is set to AC) when the POLARITY INVERT button is released. Is opposite collector supply polarity (negative-going in AC) when either the POLARITY INVERT button is pressed or the Lead Selector switch is set to BASE GROUNDED. If Lead Selector switch is set to BASE GROUNDED, POLARITY INVERT button has no effect on steps and offset polarity.
Step Families Repetitive families of characteristic curves generated with REP STEP FAMILY button pressed. Single family of characteristic curves generated each time SINGLE STEP FAMILY button is pressed.
Number of Steps Ranges from 1 to 10 as selected by the NUMBER OF STEPS switch. For zero steps, press SINGLE STEP FAMILY button.

Display Amplifiers

Display Accuracies (% of Highest On-Screen Value)
Display magnified (DISPLAY OFFSET Selector switch set to either VERT X10 or HORIZ X10) and offset between Display Unmagnified
Normal and DC Collector Supply Modes 100 and 40 divisions 35 and 15 divisions 10 and 0 divisions
Vertical Collector Current 2% 3% 4% 3%
External Vertical (Through Interface) 2% 3% 4% 3%
Horizontal Collector Volts 2% 3% 4% 3%
Horizontal Base Volts 2% 3% 4% 3%
External Horizontal (Through Interface) 2% 3% 4% 3%
Leakage Collector Supply Mode
Vertical Emitter Current (VERTICAL Switch set between 10 nA and 2 mA) 2% ±1 nA 3% ±1 nA 4% ±1 nA 3% ±1 nA
Vertical Emitter Current (VERTICAL Switch set to 5 nA, 2 nA or 1 nA) Not Applicable 5% ±1 nA
Horizontal Collector or Base Volts VERTICAL switch set to:
1 μA or more 2% 3% 4% 5%
100 nA, 10 nA or 1 nA Not Applicable 3% plus 0.025 V for each vertical division of deflection on the CRT
500 nA, 50 nA or 5 nA Not Applicable 3% plus 0.125 V for each vertical division of deflection on the CRT
200 nA, 20 nA or 2 nA Not Applicable 3% plus 0.050 V for each vertical division of deflection of the CRT
Step Generator Display
Vertical Step Generator 3% 4% 5% 4%
Horizontal Step Generator 3% 4% 5% 4%
Deflection Factors
Vertical Collector Current 1 μA/division to 2 A/division in 1-2-5 sequence.
Vertical Emitter Current 1 nA/division to 2 rnA/division in 1-2-5 sequence.
Vertical Step Generator 1 step/division.
Horizontal Collector Volts 50 mV /division to 200 V /division in 1-2-5 sequence
Horizontal Base Volts 50 mV /division to 2 V/division in 1-2·5 sequence.
Horizontal Base Volts Input Impedance At least 100 MΩ with HORIZONTAL switch set to 50 mV, 100 mV and 200 mV BASE; 1 MΩ within 2% with switch set to .5 V, 1 V and 2V.
Horizontal Step Generator 1 step/division
Maximum Displayed Noise 1% or less, or the following depending on setting of MAX PEAK VOLTS switch:
15 75 350 1500
Vertical Collector 1 μA 1 μA 2 μA 5 μA
Vertical Emitter 1 nA 1 nA 2 nA 5 nA
Horizontal Collector 5 mV 5 mV 20 mV 200 mV
Horizontal Base 5 mV 5 mV 5 mV 5 mV
Calibration Check With DISPLAY OFFSET Selector switch set to NORM (OFF), spot is deflected 10 divisions both vertically and horizontally within 1.5% whenever the CAL button is pressed.

With DISPLAY OFFSET Selector switch set to X10 MAGNIFIER (either axis) the calibration spot is within 0.5% of zero spot (previously set to CRT graticule center) when CAL button is pressed.

Vertical and Horizontal Position Controls Coarse positioning in 5 division increments within 0.1 division; continuous fine positioning over at least 5 divisions for each coarse position.
Display Offset Vertical or Horizontal offset of display centerline value up to 10 divisions in 21 half division steps.
Display Positioning Accuracy Using POLARITY Switch Spot positioning with change in POLARITY switch setting (using AC position as reference), within 0.1 division of:
Vertically Horizontally
AC Centered Centered
+(NPN) -5 divisions -5 divisions
-(PNP) +5 divisions +5 divisions

CRT and Readout

CRT
Type Electrostatic deflection.
Screen Size Calibrated area of 10 divisions by 10 divisions; 12 usable divisions horizontally ( 1 division equals 1cm).
Typical Accelerating Potential 4000 V
Readouts Automatic digitally lighted display. Readout is automatically blanked if readings would be outside the available ranges or would give erroneous display.
PER VERT DIV 1 nA to 20 A calculated from VERTICAL switch setting, DISPLAY OFFSET Selector switch setting and MODE switch setting (or X 10 Vertical Interface Input).
PER HORZ DIV 5 mV to 200 V calculated from HORIZONTAL switch setting and DISPLAY OFFSET Selector switch setting.
PER STEPS 5 nA to 2A and 5 mV to 20 V calculated from AMPLITUDE switch setting and .1X STEP MULT button position (or X10 Step Interface Input).
ß or Gm PER DIV
1 µ to 500 k calculated from VERTICAL switch setting, DISPLAY OFFSET Selector switch setting, AMPLITUDE switch setting, .1 X STEP MULT button position, X10 Vertical Interface Input and X10 Step Interface Input.

Power Requirements

Power Connection This instrument is designed for operation from power source with its neutral at or near ground (earth) potential. It is not intended for operation from two phases of multi-phase system, or across legs of single-phase, three wire system. It is provided with a three-wire power cord with three-terminal polarized plug for connection to the power source. Third wire is directly connected to instrument frame, and is intended to ground the instrument to protect operating personnel, as recommended by national and international safety codes.
Line Voltage Range Setting 115 VAC 230 VAC
Low 90 V to 110 V 180 V to 220 V
Medium 104 V to 126 V 208 V to 252 V
High 112V to 136 V 224 V to 272 V
Line Frequency Range 48 to 66 Hz
Maximum Power Consumption at 115 VAC, 60 Hz 305 W, 3.2 A

Enviromental Characteristics

Temperature
Nonoperating -40°C to +65°C
Useful Operation 0°C to +50°C
Specified Operation +10°C to +40°C
Altitude
Nonoperating To 50,000 feet
Specified Operation To 10,000 feet
Vibration
Operating 15 minutes along each axis at 0.015 inch with frequency varied from 10-50-10 c/s in 1-minute cycles. Three minutes at any resonant point or at 50 c/s.
Shock
Nonoperating 30 g's, 1/2 sine, 11 ms duration, 1 shock per axis. Total of 6 shocks
Transportation 12 inch package drop. Qualified under the National Safe Transit Committee test procedure 1A.

Mechanical Characteristics

Dimensions
Height ≈15 inches
Width ≈11 3/4 inches
Depth ≈23 1/4 inches
Weight ≈69 lbs.
Finish
Front Panel (Type 576 and Standard Test Fixture) Anodized Aluminum
Cabinet Blue vinyl painted aluminum
Trim and Rear Panel Satin finished chrome

Notes:
¹ Collector Supply Maximum Continuous Peak Current Operating Time vs Duty Cycle and Ambient Temperature. With the PEAK POWER WATTS at 50 only, the following limitations apply: Maximum continuous operating time at rated current (100% duty cycle) into a short circuit is 20 minutes at 25° C ambient, or 10 minutes at 40° C ambient. Alternatively, duty cycle may be limited to 50% at 25° C ambient or 25% at 40° C ambient. (A normal family of curves for a transistor will produce a duty cycle effect to 50% or less even if operated continuously.) Over dissipation of the collector supply will temporarily shut it off and turn on the yellow COLLECTOR SUPPLY VOLTAGE DISABLED light. No damage will result.
² Continuous DC Output vs Time, Temperature and Duty Cycle. 2 A continuous DC output can be achieved for an unlimited period up to 30° C ambient. Between 30° C and 40° C ambient, 2 A continuous DC operation should be limited to 15 minutes or limited to a 50% duty cycle or less. A family of steps (such as 10 steps at 200 mA per step) will automatically reduce the duty cycle to 50% even if generated continuously. Exceeding the rating will temporarily shut off power to the entire instrument but no damage will result.