P6019
The P6019 is a 125 turn split-core AC current probe introduced in 1966 together with passive termination and the Type 134 amplifier.
The P6020, introduced at the same time, is a 50 turn, higher frequency companion introduced at the same time.
The P6019 and P6020 have no internal frequency compensation components, contrary to their 1969 successors P6021 and P6022.
Specifications
P6019 + 011-0078-00 passive termination | P6019 + 134 | P6020 + 011-0079-00 passive termination | P6020 + 134 | |
---|---|---|---|---|
Sensitivity | 2 mA/mV or 10 mA/mV | 1 mA/Div to 1 A/Div (scope @ 50 mV/Div) | 1 mA/mV or 10 mA/mV | 1 mA/Div to 1 A/Div (scope @ 50 mV/Div) |
Bandwidth (-3 dB) | 60 MHz @ 10 mA/mV | > 40 MHz | 200 MHz @ 10 mA/mV | > 70 MHz |
LF limit (-3 dB) | 450 Hz @ 2 mA/mV, 120 Hz @ 10 mA/mV | < 12 Hz | 6500 Hz @ 1 mA/mV, 775 Hz @ 10 mA/mV | < 50 Hz |
Maximum current | 15 Ap-p | 6 Ap-p |
Links
Pictures
-
011-0079-00 passive termination for P6020
-
011-0079-00 passive termination for P6020