TG501
The Tektronix TG501 is a time mark generator plug-in for the TM500 system.
Note the output must be terminated into 50 Ω or there may be no signal at all.
There is also a TG501A.
Specifications
- Markers: 5 s to 1 ns in a 1,2,5 sequence
- Marker Amplitude: ≥1V peak on 5s through 10 ns; ≥750 mVp-p on 5 ns and 2 ns; ≥ 200 mVp-pV on 1 ns - All values into 50 Ohm
- Trigger Output Signal: Slaved to marker output from 5 s through 100 ns . Remains at 100 ns for all faster markers
Internal Reference | Standard - 1MHZ | Option:1 - 5 MHz |
---|---|---|
Stability | 1E-5 | 5E-7 |
Long-Term Drift | 1E-5 per month | 1E-7 per month |
Setability | 1E-7 | 5E-9 |
- External Reference Input: 1 MHz; 5 MHz; 10 MHz; TTL compatible (Internally hard wired so output frequencies counted down to 1 MHz)
- Error Readout Range: to ±7,5%
- Error Readout accuracy: One digit (±0,1%-points)
- Net Weight: 1 kg
- Dimensions: 66,8 mm (w/B) x 285,3 mm (d/T) 125,9 mm (h/H)
Repair Issues
- Check for +15V and adjust if not between 14,3 V and 15,7 V
- If +5V is not in between 4,94 V and 5,46 V check U626 (LM741 OpAmp). The old OpAmps tend to fail.
- A lot of ICs are placed in non gold plated IC sockets. Contact problems can occure. Removing and inserting ICs helps.
- Mounting of A1 board to frame should be done with washers in between, especially near the rear connector. Shorts to conductor tracks can occure otherwise.