PG506/Repairs

From TekWiki
Revision as of 06:27, 22 April 2024 by Qfissler (talk | contribs) (SRF2320 - more info)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigation Jump to search

Replacement front plate

Repair hints

If your PG506 fails to show a display of the Variable Value on the display, check the wave forms at U400 and U470. In case of failure in the "STD. AMPLITUDE_Variable" mode, where the display and HIGH_LOW indicators do not operate; check for these wave forms at U470 and U400. This assumes that the display momentarily shows "8.8" when the variable switch is pressed "in". If the display does not show "8.8", then there are other issues related to the display and not the DVM circuit.

U400 = SN74LS00N Link to datasheet here: https://www.ti.com/lit/gpn/SN74LS00
U470 = TL710CP or equivalent

Here is a link to photos of these wave forms. There are two photos of a legal pad and hand notations which are an index key to associate the photos to the various pins. Photos of U400 were taken at each pin and with the variable in Full CCW. ZERO and Full CW positions. With few exceptions, there are 3 photos per pin. Exceptions are Vcc, Ground and a few others. Please refer to the proper PG506 Service Manual for more details. These photos were taken on the RELAY Equipped Version of PG506 (Not "A")S/N is B040000 and above.

PG 506 U400 AND U470 Wave forms https://www.icloud.com/sharedalbum/#B0e532ODWGRrNkO

If you have questions contact me at >mlynch002@gmail.com< or >mlynch003@yahoo.com<.

PG506 Settings:

STD AMPLITUDE: SELECTED
AMPLITUDE VARIABLE: OUT
AMPLITUDE SWITCH: 1V

WAVE FORMS Photographed on TDS460A. See photos for scope settings as these can vary from photo to photo.

During this repair, I found these parts defective. The repairs were much more than a non-functional display and some of the components below were related to those additional problems. The display was the last thing to get working.

In the course of the repairs, I found these components bad in the unit.

Cct ref Tek Part Type Case Generic type Notes
Q280 151-0280-00 PNP BJT TO-39 2N4931, 200 mA, 250 V, 1W
Q290 151-0280-00 PNP BJT TO-39 2N4931, 200 mA, 250 V, 1W
Q745 151-0446-00 NPN BJT Motorola SRF2320, Is this the 2N2320? (2N2320 is superseded by 2N834 which is TO-18)
Q435 151-0223-00 NPN BJT TO-92-3 2N5769, 200 mA, 40 V, 350 mW
Q535 151-0223-00 NPN BJT TO-92-3 2N5769, 200 mA, 40 V, 350 mW
U460 156-0067-00 OP amp 8-DIP 741CN There are better performing modern drop-in replacements for the 741
U200 156-0067-00 OP amp 8-DIP 741CN There are better performing modern drop-in replacements for the 741
U240 156-0067-00 OP amp 8-DIP 741CN There are better performing modern drop-in replacements for the 741
U610 156-0039-00 Dual J-K Flip-Flop 14-DIP 7473PC
U480 156-0399-00 Opto Coupler 6-DIP MOC102 / 4N27
U255 156-0399-00 Opto Coupler 6-DIP MOC102 / 4N27
CR290 152-0141-02 Small signal diode DO-35 1N4152R, 30 V , 150 mA
CR291 152-0141-02 Small signal diode DO-35 1N4152R, 30 V , 150 mA

If your PG506 only outputs slightly more than half of the intended voltage in standard amplitude mode, check the DC-AC (S2020) switch on the A2 board for contact issues by slightly wiggling it in a circular motion. If the switch was not moved for a long time, the switch contacts might need to be cleaned.