468: Difference between revisions

From TekWiki
Jump to navigation Jump to search
Line 43: Line 43:
468 front.JPG|front
468 front.JPG|front
468 top.JPG|top
468 top.JPG|top
Top Digital.JPG|Top Digital
Top Digital.JPG|Storage & Display A16
468 right.JPG|right
468 right.JPG|right
468 left.JPG|left
468 left.JPG|left
Line 49: Line 49:
468 back.JPG|rear
468 back.JPG|rear
Storage Boards.JPG|Storage Boards
Storage Boards.JPG|Storage Boards
CPU Board.JPG|CPU Board
CPU Board.JPG|CPU Board A21
</gallery>
</gallery>



Revision as of 00:24, 23 March 2021

{{{manufacturer}}} 
10/100 MHz Digital Storage Scope
Tek 468

Produced from 1980 to 1985

Manuals
(All manuals in PDF format unless noted otherwise)
Manuals – Specifications – Links – Pictures

The Tektronix 468 is a digital/analog scope, essentially a 465B with an added 25 MS/s 8 bit digitizer.

Specifications

  • Digital mode: Sampling rate 25 MS/s, bandwidth 10 MHz (sine)/2.5 MHz (pulse), resolution 8 bit
  • Analog mode: 100 MHz bandwidth
  • GPIB option
  • please add more ...

Repair issues

Many 468s used Mostek MKB36000 masked ROMs that can suffer from "ROM rot". It may be necessary to replace the 160-0459-01 and 160-0759-01 ROMs with 2764 EPROMs through a 2364/2764 adapter. Another option might be using the MCM68766 EPROM.

ROM images


A16 STORAGE-DISPLAY Circuit Board Replacement ERROR in SERVICE Manual Schematic and Parts list (See Repairs)

Links

Pictures