The interface connections to the accessory modules are made through the same style of blue robust Amphenol 26-xxx-xx connectors as used with the plug-ins on the 500-series oscilloscope line. | The interface connections to the accessory modules are made through the same style of blue robust Amphenol 26-xxx-xx connectors as used with the plug-ins on the 500-series oscilloscope line. | ||
The collector measurement range is from 100 nA/Div to 2 A/Div. | The collector measurement range is from 100 nA/Div to 2 A/Div. | ||
Base voltage can be swept from zero to 1500 V in four ranges. | |||
The three higher ranges require the [[337-1194-02|safety cover]] to be closed. | |||
Because so many used 576s do not come with safety covers, most users defeat the safety interlock mechanically or electrically. | |||
If this is done, the operator should be especially cautious as the 576 is capable of producing lethal voltages and currents, up to 1500 V at up to 100 mA continuous. | |||
A calibration fixture, the [[067-0599-00]], was produced for the 576. | A calibration fixture, the [[067-0599-00]], was produced for the 576. |