576/Repairs: Difference between revisions

Line 63: Line 63:
Lost +12.5 and therefore +225 as well. Also affected were 5V and 4.5V which were both about half the required voltage.
Lost +12.5 and therefore +225 as well. Also affected were 5V and 4.5V which were both about half the required voltage.


==Defects found without any noted symptoms==
==Defects found without any noted symptoms. . . Could these be part or all of the cause for failures previously noted?==
==NOTE: Just because I did not see "symptoms" does not mean that they were not there, just that I was not experienced enough to find any such symptoms.==  
Found D507-D708 9V Zener VR TC shorted (continuity both directions) on both Step Gen and Display Board.
Found D507-D708 9V Zener VR TC shorted (continuity both directions) on both Step Gen and Display Board.
Tek's recommended replacement is a [[1N936]], available at [https://www.mouser.com Mouser Electronics].
Tek's recommended replacement is a [[1N936]], available at [https://www.mouser.com Mouser Electronics].
Found shorted D83 -  1N4152.
D507 was replaced with a TC Diode, D708 was replaced with a standard 9V ZENER (not TC, did not have the proper part on hand),
Replaced  Symptoms not noted, but seems to have less distortion in the display.
 
RESULT: -75V drifts badly as the machine is turned on and heats up (High @ -73.5V when turned on and drifting slowly to -75.05V when warmed up.).  
   
   
NOTE: All electrolytic caps which were previously replaced, (except for one cap) tested as high ESR to the point of almost being shorts to ground.  
Found shorted D83 -  1N4152 not aware of symptoms related to this defect.
Undoubtedly loading they power supply and not filtering much at all.
   
Replaced all remaining electrolytics that could be found on the Right side.  
Replaced D708 with proper 9V TC ZENER, -75V Supply is now rock solid @-75.015V, no "drift" as the machine heats up.
NOTE: All electrolytic caps which were previously replaced, (except for one cap) tested as either high ESR or to the point or being shorts to ground.  
Undoubtedly loading they power supply and not filtering much at all.
 
Replaced all remaining electrolytics that could be found on the Right side and Power supply Filter caps.  
Voltages in all supplies became cleaner and most of the voltages increased as well.  
Voltages in all supplies became cleaner and most of the voltages increased as well.  
Power supply is now well within the normal operating range.  
Power supply is now well within the normal operating range.  
Type 576 electrolytic caps should be your primary suspect when repairing or adjusting these machines.       
Type 576 electrolytic caps should be your primary suspect when repairing or adjusting these machines.       