067-0916-00

From TekWiki
Jump to navigation Jump to search
Tektronix 067-0916-00
video amplitude calibration fixture
Tektronix 067-0916-00

Produced from (?) to (?)

Manuals
(All manuals in PDF format unless noted otherwise)
Manuals – Specifications – Links – Pictures

The Tektronix 067-0916-00 is a video amplitude calibration fixture plug-in for the TM500 system.

It is used for measuring and calibrating luminance and chrominance waveform characteristics of video signal generators and waveform monitors.

Key Specifications

  • please add

The 067-0916-00 (VAC) generates a precise square wave from its low impedance output with a resolution of 0.1 mV and an accuracy of 0.05% from 0 to 999.9 mV. Color standards NTSC, PAL, PAL-M, and SECAM along with other signal parameters such as half or full amplitude and many others may be selected.

The 067-0916-00 (VAC) came with a subcarrier harmonic rejection filter (015-0407-00), 0.6% attenuator (011-0134-00), and precision terminator (011-0102-01).

Links

Pictures