Line 18: Line 18:
or [[A6303]] probe (52.5 A continuous).
or [[A6303]] probe (52.5 A continuous).


The current probe input is an [[Amphenol 165-12 connector]]. It contains pins for hall element and transformer connections, lock detection, probe degaussing, and probe type encoding. The 11A16 detects the current probe type and scales the display appropriately.
The current probe input is an [[Amphenol 165-12 connector]].
It contains pins for hall element and transformer connections,
lock detection, probe degaussing, and probe type encoding.
The 11A16 detects the current probe type and scales the display appropriately.
 
The 11A16 uses one [[M377]] amplifier chip per channel, U220 and U260.