32,948
edits
(transitory commit, not complete) |
(transitory commit, not complete) |
||
Line 36: | Line 36: | ||
except for 12 additional constants required by the 11A52. | except for 12 additional constants required by the 11A52. | ||
'''Note''': The 12 calibration constants for 11A52 high frequency transient response cannot be set using an 11300 series mainframe. | '''Note''': The 12 calibration constants for 11A52 high frequency transient response | ||
This is apparently due to a failure of the | cannot be set using an 11300 series mainframe. | ||
in spite of the fact that the two groups were contiguous and had the same manager. | This is apparently due to a failure of the plug-in design group and the 11300 series design group | ||
to communicate in spite of the fact that the two groups were contiguous and had the same manager. | |||
When given a choice, do not use an 11401 main frame for setting the 11A52 high frequency calibration constants | When given a choice, do not use an 11401 main frame for setting the 11A52 high frequency calibration constants | ||
as the 11401 bandwidth is not quite adequate to adjust the 11A52 properly, although you can probably come pretty close. | as the 11401 bandwidth is not quite adequate to adjust the 11A52 properly, | ||
although you can probably come pretty close. | |||
The DSA601 mainframe can be used as well as the 11402 and 11403. | The DSA601 mainframe can be used as well as the 11402 and 11403. | ||
For all | |||
For all plug-ins, you must enter a serial number into the NVRAM using an 11000 series mainframe, | |||
a personal computer and either an RS232C cable or GPIB cable connected between them. | a personal computer and either an RS232C cable or GPIB cable connected between them. | ||