==CRT Circuit Troubles==
==CRT Circuit Troubles==
The 465 "Family" of scopes has been known to have issues with "no display" or a "shrunken" screen trace.  This includes 465,465B,466,468 and 475 scopes.  The schematic for each series is almost identical but each one uses different and component ID numbers in their scematic.   
==The 465 "Family" of scopes has been known to have issues with "no display" or a "shrunken" screen trace.  This includes 465,465B,466,468 and 475 scopes.  The schematic for each series is almost identical but each one uses different and component ID numbers in their scematic.   
       1.  This may be attributed to a failure of the HV transformer and/or the HV Multiplier.   
       1.  This may be attributed to a failure of the HV transformer and/or the HV Multiplier.   
           a.  These parts are generally shared across the entire family of instruments.   
           a.  These parts are generally shared across the entire family of instruments.   
Line 53: Line 53:
       11.  Re-Test all voltages in the CRT circuit and confirm that they generally agree with the schematic.
       11.  Re-Test all voltages in the CRT circuit and confirm that they generally agree with the schematic.
       12.  Retest for the same waveforms (Step 6 and 8) as before, these should be more close to the ones in the diagrams.   
       12.  Retest for the same waveforms (Step 6 and 8) as before, these should be more close to the ones in the diagrams.   
       13.  Locate a known good HV Multiplier and replace as needed.
       13.  Locate a known good HV Multiplier and replace as needed.==