Telequipment Oscilloscope Calibrator C3: Difference between revisions

Jump to navigation Jump to search
m
no edit summary
(added page and specs)
 
mNo edit summary
Line 26: Line 26:
Output impedance varies from 5 kΩ at 50 V and 50 mV down to 75Ω at ≤ 0.5 V and 0.5 mV. }}
Output impedance varies from 5 kΩ at 50 V and 50 mV down to 75Ω at ≤ 0.5 V and 0.5 mV. }}


{{Spec | Fast Rise Signal | Square waveform switched at 100kHz, 1MHz, 1OMHz.<br />
{{Spec | Fast Rise Signal | Square waveform switched at 100kHz, 1MHz, 10MHz.<br />
Risetime ≤ 1 nsec.<br />
Risetime ≤ 1 nsec.<br />
Output 0.2 V maximum into terminated 50 Ω load. }}}
Output 0.2 V maximum into terminated 50 Ω load. }}}
Line 45: Line 45:
Amplitude 2.5 volts approximately.<br />
Amplitude 2.5 volts approximately.<br />
Output impedance 1 kΩ approximately. }}}
Output impedance 1 kΩ approximately. }}}
{{Spec | Ext Clock | External clock frequencies of 2V peak to peak amplitude may be fed to the 1O MHz Auxiliary Output socket. }}}
{{Spec | Ext Clock | External clock frequencies of 2V peak to peak amplitude may be fed to the 10 MHz Auxiliary Output socket. }}}
{{Spec | Power Supply | 100, 110, 120, 200, 220, 240 V, 48 to 440 Hz, 13 VA. }}}
{{Spec | Power Supply | 100, 110, 120, 200, 220, 240 V, 48 to 440 Hz, 13 VA. }}}
{{Spec | Size | Height 107 mm<br />
{{Spec | Size | Height 107 mm<br />