475/Repairs: Difference between revisions

Jump to navigation Jump to search
no edit summary
No edit summary
No edit summary
Line 36: Line 36:


More info to follow.
More info to follow.
==High Voltage CRT Circuit==
As with most 400 series scopes, the high voltage circuits decay over time. It will mainly be the non-ceramic axial capacitors that go bad.
This is indicated by a crackling constant arcing noise that can be triggered with the varying of intensity control knob.
Replacement capacitors are available on Digikey and Mouser. However, they will usually be available only in large bulk amounts so sourcing components may carry difficulties.


==General Troubleshooting==
==General Troubleshooting==
130

edits

Navigation menu