Page values for "DC501"

"_pageData" values

1 row is stored for this page
FieldField typeValue
_creationDateDatetime2011-10-24 3:10:39 AM
_modificationDateDatetime2023-11-19 3:40:27 PM
_categoriesList of String, delimiter: |Introduced in 1972 TM500 counter plugins
_isRedirectBooleanNo
_pageIDInteger1,917
_pageNamePageDC501
_pageTitleString

DC501

_pageNamespaceInteger0

"Products" values

1 row is stored for this page
FieldField typeValue
ManufacturerPageTektronix
ModelStringDC501
ClassStringPlug-in
DescriptionWikitext string

basic 100 MHz frequency counter

SeriesPageTM500 system
IntroducedText1972
DesignersList of Page, delimiter: ;Jim Geddes
DiscontinuedText1980

"Specifications" values

7 rows are stored for this page
FieldField typeValue
ParameterStringInput frequency
ValueWikitext

10 Hz to 100 MHz

FieldField typeValue
ParameterStringGate time
ValueWikitext

10 ms to 10 s in decade steps, plus Manual (totalize count of up to 107 events)

FieldField typeValue
ParameterStringResolution
ValueWikitext

7 digits

FieldField typeValue
ParameterStringDisplay
ValueWikitext

7-digit LED, overflow indicator

FieldField typeValue
ParameterStringInput impedance
ValueWikitext

1 MΩ // 20 pF, ×5/×10/×50 attenuator (front) or 50 Ω // 20 pF (rear)

FieldField typeValue
ParameterStringTime base stability
ValueWikitext

≤1×10-5; Opt. 1, ≤5×10-7 (0°C to +50°C after ½h warm-up)

FieldField typeValue
ParameterStringTime base drift
ValueWikitext

≤1×10-5; Opt. 1, ≤1×10-7 (per month)