Page values for "DC503"

Jump to navigation Jump to search

"_pageData" values

1 row is stored for this page
FieldField typeValue
_creationDateDatetime2011-10-24 3:12:14 AM
_modificationDateDatetime2023-12-06 5:12:35 PM
_categoriesList of String, delimiter: |Introduced in 1972 TM500 counter plugins
_isRedirectBooleanNo
_pageIDInteger1,918
_pageNamePageDC503
_pageTitleString

DC503

_pageNamespaceInteger0

"Products" values

1 row is stored for this page
FieldField typeValue
ManufacturerPageTektronix
ModelStringDC503
ClassStringPlug-in
DescriptionWikitext string

100 MHz, 7-digit frequency counter

SeriesPageTM500 system
IntroducedText1972
DesignersList of Page, delimiter: ;Roland Crop Jim Geddes
DiscontinuedText1981

"Specifications" values

12 rows are stored for this page
FieldField typeValue
ParameterStringResolution
ValueWikitext

7 digits

FieldField typeValue
ParameterStringFrequency Range
ValueWikitext

0.1 Hz (AC coupled: 10 Hz) to 100 MHz

FieldField typeValue
ParameterStringImpedance
ValueWikitext

1 MΩ paralleled by approx. 20 pF

FieldField typeValue
ParameterStringSensitivity
ValueWikitext

300 mVp-p sinewave times attenuation to 100 MHz

FieldField typeValue
ParameterStringMax Input Voltage, 1× Att.
ValueWikitext

50 V @ ≤2 kHz, derate –20 dB/decade to 10 MHz; 10 V @ 10 kHz to 25 MHz, derate –20 dB/decade to 50 MHz; 5 V @ 50 MHz to 100 MHz

FieldField typeValue
ParameterStringMax Input Voltage, 10×/100× Att.
ValueWikitext

500 V @ ≤2 MHz, derate –20 dB/decade to 100 MHz

FieldField typeValue
ParameterStringFrequency
ValueWikitext

1 MHz ± 1 × 10-7 at cal

FieldField typeValue
ParameterStringError
ValueWikitext

≤ ±1 × 10-5 from 0 °C to +50 °C

FieldField typeValue
ParameterStringAging
ValueWikitext

≤ ±1 × 10-5 per month

FieldField typeValue
ParameterStringFrequency
ValueWikitext

5 MHz ± 5 × 10-9 at cal

FieldField typeValue
ParameterStringError
ValueWikitext

≤ ±5 × 10-7 from 0 °C to +50 °C (after oven warm-up)

FieldField typeValue
ParameterStringAging
ValueWikitext

≤ ±1 × 10-7 per month