Page values for "Keithley 610C"

Jump to navigation Jump to search

"_pageData" values

1 row is stored for this page
FieldField typeValue
_creationDateDatetime2024-05-19 8:52:07 AM
_modificationDateDatetime2024-10-11 1:53:11 PM
_categoriesList of String, delimiter: |
_isRedirectBooleanNo
_pageIDInteger35,570
_pageNamePageKeithley 610C
_pageTitleString

Keithley 610C

_pageNamespaceInteger0

"Products" values

1 row is stored for this page
FieldField typeValue
ManufacturerPageKeithley
ModelString610C
ClassStringMultimeter
DescriptionWikitext string

Solid-state Electrometer

SeriesPage
IntroducedText1968
DesignersList of Page, delimiter: ;
DiscontinuedText(?)

"Specifications" values

10 rows are stored for this page
FieldField typeValue
ParameterStringVoltage
ValueWikitext

1 mV to 100 V full scale in 11 ranges (×1/×3)

FieldField typeValue
ParameterStringInput resistance (volts mode)
ValueWikitext

>1014 Ω or selectable 10 Ω to 1011 Ω in decade steps

FieldField typeValue
ParameterStringAccuracy
ValueWikitext

1% of full scale, noise 25 μV

FieldField typeValue
ParameterStringZero offset drift
ValueWikitext

<1 mV per day / 150 μV per °C

FieldField typeValue
ParameterStringCurrent
ValueWikitext

10–14 A to 0.3 A full scale in 28 ranges (×1/×3)

FieldField typeValue
ParameterStringResistance
ValueWikitext

100 Ω to 1014 Ω full scale in 25 linear ranges (×1/×3)

FieldField typeValue
ParameterStringCharge
ValueWikitext

10–13 C to 10–3 C (Ampere-seconds) full scale in 17 ranges (×1/×3)

FieldField typeValue
ParameterStringDisplay
ValueWikitext

Analog meter with mirror scale; positive, negative and center-zero display modes

FieldField typeValue
ParameterStringPower
ValueWikitext

105−125 V or 210−250 V, 50−60 Hz (some units labelled 50−1000 Hz), 10 W

FieldField typeValue
ParameterStringAccessories
ValueWikitext

  • Model 2503 Static Detector Probe (for measurements of charge on flat surfaces)
  • Model 6101A Shielded Probe (shielded low-noise cable with a needle-point probe)
  • Model 6103C Divider Probe (1000:1 voltage divider with a 3.8×1011 input resistance)
  • Model 6105 Resistivity Chamber (test fixture for measurement of surface and volume resistivity)