For repair or for building your own temperature sensor you need a PT1-100 sensor and a LEMO connector.
The LEMO connector:
- Series: S
- The diameter: 0S (S-Series, diameter 0)
- Insert configuration: 302 (2 pins)
The following LEMO configuration is known to fit:
- Connector: FFA0S302CLA (2.2 to 4.4 mm cable)
- Bend relief: GMA0B035 (3.5 to 3.9 mm cable)
As the PT-100 is connected with only 2 wires, each sensor has to be trimmed with R1001 (offset).