067-0680-00: Difference between revisions
Jump to navigation
Jump to search
No edit summary |
m (change 5000 to 5400) |
||
(4 intermediate revisions by 3 users not shown) | |||
Line 1: | Line 1: | ||
{{Plugin Sidebar | {{Plugin Sidebar | ||
|manufacturer=Tektronix | |manufacturer=Tektronix | ||
|series= | |series=5400 Series Scopes | ||
|type=067- | |type=067-0680-00 | ||
|summary= | |summary=5400 Series Mainframe Calibrator | ||
|image=067-0680-00-front.jpg | |image=067-0680-00-front.jpg | ||
|caption=Tektronix 067- | |caption=Tektronix 067-0680-00 | ||
|introduced=(?) | |introduced=(?) | ||
|discontinued=(?) | |discontinued=(?) | ||
|designers= | |designers= | ||
|manuals= | |manuals= | ||
* [[Media:062-1587-01.pdf|Tektronix 067-0680-01 Manual, | * [[Media:062-1587-01.pdf|Tektronix 067-0680-01 Manual, 2nd Rev. 1977]] (OCR) | ||
* [[Media:062-1587-00.pdf|Tektronix 067-0680-00 Manual, | * [[Media:062-1587-00.pdf|Tektronix 067-0680-00 Manual, 1st Rev.]] (OCR) | ||
}} | }} | ||
The '''Tektronix 067-0680-00''' is a plug-in used to calibrate a 5400-series mainframe's | The '''Tektronix 067-0680-00''' is a plug-in used to calibrate a 5400-series mainframe's |
Latest revision as of 21:04, 15 November 2023
The Tektronix 067-0680-00 is a plug-in used to calibrate a 5400-series mainframe's horizontal and vertical deflection factors and step response, and help align CRT geometry.
It provides a calibrated stair-step signal for deflection factor and geometry adjustment, a fast-risetime step function for checking and trimming step response, a leveling circuit for an external sinewave test signal, and a common-mode input test for both the vertical and trigger systems.
It is similar to the 067-0587-xx for 7000 series scopes.
Key Specifications
- please add