{{Instrument Sidebar | |||
|manufacturer=Tektronix | |||
|model=067-0532-01 | |||
|class=Signal generator | |||
|summary=500 MHz Constant-Amplitude Signal Generator | |||
|image=067-0532-01 front.JPG | |||
|caption=Tektronix 067-0532-01 Constant Amplitude Signal Generator | |||
|series= | |||
|introduced=1971 | |||
|discontinued=(?) | |||
|manuals= | |||
* [[Media:062-1176-00.pdf|Tektronix 067-0532-01 Manual (062-1176-00)]] | |||
}} | |||
The Tektronix '''Type 067-0532-01 Calibration Fixture''' is a constant amplitude signal generator whose output is a low-distortion sinewave. | |||
It is intended to provide a signal of known frequency and amplitude for measuring oscilloscope bandwidth. | |||
A 3 MHz sinewave is available as a reference amplitude from which a | A 3 MHz sinewave is available as a reference amplitude from which a –3 dB measurement can be made. | ||
The output connector, amplitude detector, and 50 Ω source resistor are situated at the end of a 42 inch cable. | |||
Output amplitude is regulated at the end of the cable adjacent to the [[GR-874 connector|GR-874 output connector]], thus essentially eliminating cable loss and standing-wave errors. | |||
{{BeginSpecs}} | {{BeginSpecs}} | ||
{{Spec | Frequency range | 65 MHz to 500 MHz}} | |||
{{ | {{Spec | Frequency accuracy | ±2% of indicated frequency; 3 MHz output ±5%}} | ||
{{Spec | Amplitude range | 0.5 V<sub>p-p</sub> to 4 V<sub>p-p</sub> into 50 Ω in 0.5 V<sub>p-p</sub> increments; variable between calibrated steps; max. output 4.2 V<sub>p-p</sub> }} | |||
{{Spec | | {{Spec | Amplitude accuracy | 3 MHz ±2%, 65−300 MHz ±2%, 300−500 MHz ±5% }} | ||
{{Spec | Output | 50 Ω ±1%, via [[GR-874 connector]] }} | |||
{{ | |||
{{Spec | | |||
{{Spec | Output | |||
{{EndSpecs}} | {{EndSpecs}} | ||
==Pictures== | ==Pictures== | ||
<gallery> | <gallery> | ||
067-0532-01 front.JPG | Front view | |||
067-0532-01 left.jpg | Left side | |||
067-0532-01 top.jpg | Top internal | |||
067-0532-01 bottom.jpg | Bottom internal | |||
067-0532-01 left-in.jpg | Left internal | |||
067-0532-01 right.jpg | Right internal | |||
067-0532-01 signal gen circuit.JPG | Signal generator circuit | |||
067-0532-01 line filter.JPG | Line filter board | |||
</gallery> | </gallery> | ||
[[Category:Calibration fixtures]] | |||
[[Category:Signal generators]] | [[Category:Signal generators]] |
Latest revision as of 03:52, 12 June 2024
The Tektronix Type 067-0532-01 Calibration Fixture is a constant amplitude signal generator whose output is a low-distortion sinewave.
It is intended to provide a signal of known frequency and amplitude for measuring oscilloscope bandwidth.
A 3 MHz sinewave is available as a reference amplitude from which a –3 dB measurement can be made. The output connector, amplitude detector, and 50 Ω source resistor are situated at the end of a 42 inch cable. Output amplitude is regulated at the end of the cable adjacent to the GR-874 output connector, thus essentially eliminating cable loss and standing-wave errors.
Key Specifications
Frequency range | 65 MHz to 500 MHz |
---|---|
Frequency accuracy | ±2% of indicated frequency; 3 MHz output ±5% |
Amplitude range | 0.5 Vp-p to 4 Vp-p into 50 Ω in 0.5 Vp-p increments; variable between calibrated steps; max. output 4.2 Vp-p |
Amplitude accuracy | 3 MHz ±2%, 65−300 MHz ±2%, 300−500 MHz ±5% |
Output | 50 Ω ±1%, via GR-874 connector |
Pictures
-
Front view
-
Left side
-
Top internal
-
Bottom internal
-
Left internal
-
Right internal
-
Signal generator circuit
-
Line filter board