067-0938-00: Difference between revisions

From TekWiki
Jump to navigation Jump to search
(cat)
No edit summary
 
(9 intermediate revisions by 4 users not shown)
Line 1: Line 1:
The Tektronix 067-0938-00 is a calibration fixture for measuring the distortion (spectral impurity) of an oscillator.
{{TM500
It is notch filter used for eliminating the fundamental frequency of the oscillator, so the residual harmonics can be measured.
|mfg=Tektronix
|type=067-0938-00
|function=passive notch filter
|class=calibration
|image=067-0938-00.jpg
|introduced=(?)
|discontinued=(?)
|designers=
|manuals=
* [[Media:070-2868-00.pdf|067-0938-00 Instruction Manual 070-2868-00]]
}}
It is used for measuring the distortion (spectral impurity) of an oscillator by eliminating the fundamental frequency of the oscillator (primarily the [[SG505]]), so the residual harmonics can be measured.


* [http://w140.com/067-0938-00_calibration_fixture.pdf Tektronix 067-0938-00 Manual (PDF)]
The circuit topology is a passive double-T RC notch filter with a precision 600 Ω load and a 60 dB attenuator. The module has no connections to the TM500 mainframe, requiring no power supply.


{{BeginSpecs}}
{{Spec|Notch frequency | 10 Hz, 20 Hz, 100 Hz, 1 kHz, 10 kHz, 20 kHz, 50 kHz, 100 kHz; ±2% }}
{{Spec|Notch depth| ≥100 dB at 10 Hz and 20 Hz, ≥60 dB otherwise }}
{{EndSpecs}}
Load impedance must be ≥1 MΩ ±5% // ≤75 pF including cabling and spectrum analyzer input capacitance.  Input source impedance must be 600 Ω ±5%.


[[Category:Calibration fixture]]
==Links==
* [http://www.barrytech.com/tektronix/tektm500/tek067-0938-00.html 067-0938-00 @ barrytech.com]
 
==Pictures==
<gallery>
Tek 067-0938-00 1.jpg
Tek 067-0938-00 2.jpg
Tek 067-0938-00 3.jpg
</gallery>
 
[[Category:Calibration fixtures]]

Latest revision as of 09:01, 18 February 2024

Tektronix 067-0938-00
passive notch filter
Tektronix 067-0938-00

Produced from (?) to (?)

Manuals
(All manuals in PDF format unless noted otherwise)
Manuals – Specifications – Links – Pictures

The Tektronix 067-0938-00 is a passive notch filter plug-in for the TM500 system.

It is used for measuring the distortion (spectral impurity) of an oscillator by eliminating the fundamental frequency of the oscillator (primarily the SG505), so the residual harmonics can be measured.

The circuit topology is a passive double-T RC notch filter with a precision 600 Ω load and a 60 dB attenuator. The module has no connections to the TM500 mainframe, requiring no power supply.

Key Specifications

Notch frequency 10 Hz, 20 Hz, 100 Hz, 1 kHz, 10 kHz, 20 kHz, 50 kHz, 100 kHz; ±2%
Notch depth ≥100 dB at 10 Hz and 20 Hz, ≥60 dB otherwise

Load impedance must be ≥1 MΩ ±5% // ≤75 pF including cabling and spectrum analyzer input capacitance. Input source impedance must be 600 Ω ±5%.

Links

Pictures