The Tektronix P6149 is a 10x voltage probe with a specified -3dB point of 50 MHz.
{{Probe Sidebar
The impedance at the tip is 10 megohms in parallel with 15.5 picofarads.
|manufacturer=Tektronix
The probe connects to the oscilloscope with a BNC connector,
|model=P6149
and the compensation range is 20 pF to 62 pF.
|summary=50 MHz, x10, passive probe
|image=Tek p6149 1.jpg
|caption=Tektronix P6149
|for=1 MΩ scope inputs
|introduced=1977
|discontinued=(?)
|manuals=
* [[Media:062-2766-00.pdf|Tektronix P6149 Data sheet]]
}}
The '''Tektronix P6149''' is a 50 MHz, ×10, passive, [[miniature modular probe]]
The impedance at the tip is 10 in parallel with 15.5 pF.
 
The probe connects to the oscilloscope with a BNC connector, and the compensation range is 20 pF to 62 pF.
 
The maximum input voltage is 500 V at DC, decreasing to about 25 V at 50 MHz.
The maximum input voltage is 500 V at DC, decreasing to about 25 V at 50 MHz.


* [http://w140.com/Tektronix_P6149_Datasheet.pdf Tektronix P6149 Datasheet (PDF)]
 
[[Category:1 MΩ Oscilloscope probes]]
[[Category:Miniature modular probes]]

Latest revision as of 11:39, 19 July 2022

Tektronix P6149
50 MHz, x10, passive probe
Tektronix P6149

Compatible with 1 MΩ scope inputs

Produced from 1977 to (?)

Manuals


Manuals – Specifications – Links – Pictures

The Tektronix P6149 is a 50 MHz, ×10, passive, miniature modular probe The impedance at the tip is 10 MΩ in parallel with 15.5 pF.

The probe connects to the oscilloscope with a BNC connector, and the compensation range is 20 pF to 62 pF.

The maximum input voltage is 500 V at DC, decreasing to about 25 V at 50 MHz.