P6129: Difference between revisions
Jump to navigation
Jump to search
No edit summary |
No edit summary |
||
(6 intermediate revisions by 2 users not shown) | |||
Line 1: | Line 1: | ||
* [[Media:070-8547-05.pdf|Tektronix P6129B Manual | {{Probe Sidebar | ||
|manufacturer=Tektronix | |||
|model=P6129 | |||
|summary=100 MHz, ×1/×10 passive probe | |||
|image=Tek p6129b 1.jpg | |||
|caption=P6129B | |||
|for=1 MΩ scope inputs | |||
|introduced=1991 | |||
|discontinued=(?) | |||
|manuals= | |||
* [[Media:070-8279-00.pdf|Tektronix P6129 Manual]] ''needed'' | |||
* [[Media:070-8547-05.pdf|Tektronix P6129B Manual]] | |||
}} | |||
The '''Tektronix P6129''' is a passive 100 MHz ×1/×10 [[miniature modular probe]], with a [[BNC connector]] that has a [[readout pin]]. | |||
{{BeginSpecs}} | |||
{{Spec|Attenuation | ×1/×10 ±1.35% at DC }} | |||
{{Spec|Input Impedance | ×1: 1 MΩ ±1.0% // 90–115 pF; ×10: 10 MΩ ±1.3% // 18.5 pF (17.5–19.5 pF) }} | |||
{{Spec|Compensation Range | 15–35 pF }} | |||
{{Spec|Bandwidth | ×1: DC–10 MHz; ×10: DC–100 MHz }} | |||
{{Spec|Maximum Input Voltage | 420 V (DC + peak AC) nondestructive }} | |||
{{EndSpecs}} | |||
==Pictures== | |||
<gallery> | <gallery> | ||
Tek p6129b 1.jpg | Tek p6129b 1.jpg | ||
Line 6: | Line 28: | ||
Tek p6129b 3.jpg | Tek p6129b 3.jpg | ||
Tek p6129b 4.jpg | Tek p6129b 4.jpg | ||
Tek p6129 in package 1.jpg | |||
Tek p6129 in package 2.jpg | |||
</gallery> | </gallery> | ||
[[Category:1 MΩ Oscilloscope probes]] | |||
[[Category:Miniature modular probes]] |
Latest revision as of 04:10, 18 July 2024
The Tektronix P6129 is a passive 100 MHz ×1/×10 miniature modular probe, with a BNC connector that has a readout pin.
Key Specifications
Attenuation | ×1/×10 ±1.35% at DC |
---|---|
Input Impedance | ×1: 1 MΩ ±1.0% // 90–115 pF; ×10: 10 MΩ ±1.3% // 18.5 pF (17.5–19.5 pF) |
Compensation Range | 15–35 pF |
Bandwidth | ×1: DC–10 MHz; ×10: DC–100 MHz |
Maximum Input Voltage | 420 V (DC + peak AC) nondestructive |