DM501A/Repairs: Difference between revisions
(Created page with "For repair or for building your own temperature sensor you need a PT1-100 sensor and a LEMO connector. The LEMO connector: *Series: S *The diameter: 0S (S-Series, diameter 0)...") |
m (501A from DM501/Repairs) |
||
(2 intermediate revisions by 2 users not shown) | |||
Line 1: | Line 1: | ||
For repair or for building your own temperature sensor you need a | For repair or for building your own temperature sensor you need a PT-100 sensor and a LEMO connector. | ||
The LEMO connector: | The LEMO connector: | ||
Line 12: | Line 12: | ||
As the PT-100 is connected with only 2 wires, each sensor has to be trimmed with R1001 (offset). | As the PT-100 is connected with only 2 wires, each sensor has to be trimmed with R1001 (offset). | ||
==Switch Repair== | |||
The contact bar in switches can break. Replacing the contact bar and contacts can be done without desoldering the switches. | |||
Refer to the the pictures and the [[260-1971-00/Repairs| repair]] section of switch assembly [[260-1971-00]]. | |||
==R10 ([[307-1015-00]] input voltage divider) Repair== | |||
Fault pattern: V<sub>DC</sub>, V<sub>AC</sub> and resistance (exept the lowest range) are 2% to low. | |||
Fault: R10. Some resistors are to low (900 kΩ and 90 kΩ). The "stuff" inside has become liquid. The liquid has a low resistance. | |||
Solution: Open R10. Clean R10 inside with isopropanol and in a ultrasonic bath. | |||
==See Also== | |||
* [[Pushbutton switches]] | |||
==Pictures== | |||
<gallery> | |||
Tek-DM501A-rep-switch.jpg| Replace the push button plunger and spring contacts on a DM501A | |||
Tek-1260-1971-00-rep4.jpg| Defect push button plunger on top and new one below | |||
</gallery> | |||
[[Category:Instrument repair reports]] |
Latest revision as of 11:17, 5 July 2024
For repair or for building your own temperature sensor you need a PT-100 sensor and a LEMO connector.
The LEMO connector:
- Series: S
- The diameter: 0S (S-Series, diameter 0)
- Insert configuration: 302 (2 pins)
The following LEMO configuration is known to fit:
- Connector: FFA0S302CLA (2.2 to 4.4 mm cable)
- Bend relief: GMA0B035 (3.5 to 3.9 mm cable)
As the PT-100 is connected with only 2 wires, each sensor has to be trimmed with R1001 (offset).
Switch Repair
The contact bar in switches can break. Replacing the contact bar and contacts can be done without desoldering the switches. Refer to the the pictures and the repair section of switch assembly 260-1971-00.
R10 (307-1015-00 input voltage divider) Repair
Fault pattern: VDC, VAC and resistance (exept the lowest range) are 2% to low.
Fault: R10. Some resistors are to low (900 kΩ and 90 kΩ). The "stuff" inside has become liquid. The liquid has a low resistance.
Solution: Open R10. Clean R10 inside with isopropanol and in a ultrasonic bath.
See Also
Pictures
-
Replace the push button plunger and spring contacts on a DM501A
-
Defect push button plunger on top and new one below